9
Low Alarm Temperature Setting: Press the Low Alarm button once to enter the Low Alarm temperature
setting mode. LO-SET will display on the LCD screen. Use the Up and Down arrows to adjust the
temperature accordingly. Then press the LOW ALARM button to confirm. The LCD screen will return to
the current temperature.
NOTE: The TMD can be locked to avoid accidental adjustments. By default, the TMD is unlocked.
Long press ºC/ºF and keys simultaneously for 6 seconds to lock all the alarm features. LOCK will
display on the screen when the unit is successfully locked. When the TMD is locked, the alarm features
and functions are inaccessible, except for silencing beeping sound.
If the temperature rises above or falls below the set temperature range, the alarm will emit a beeping
sound, and the HI-ALARM/LO-ALARM indicator on the display will blink until the temperature is back in
range. To silence the alarm, press any button. The beeping sound lasts for 60 seconds, but the high and
low indicators will remain on display even when the unit gets back in range. Press the LOW ALARM and
RESET keys simultaneously to clear the LO-ALARM/HI-ALARM display indicators. The alarm will beep
every 30 minutes if the same alarm event exists, even if the user has silenced the alarm. The alarm will
NOT clear if an alarm event still exists.
More details about the features and usage of the TMD can be found in the User Manual accompanying
the unit.
This appliance is equipped with a 3/8” probe port and plug, which can accept additional monitoring
devices. It is recommended that you position the additional temperature monitoring device on the middle
shelf close to the TMD sensor.
Door Alarm
If the door has been left open for more than 60 seconds, the alarm will sound. Pressing the POWER key
once or closing the door can switch off the audible alarm.
Door Lock
Your unit is provided with a lock and key combination. The keys are located inside the plastic bag that
contains this User Manual. Insert the key into the lock and turn it counterclockwise to unlock the door.
To lock the door, follow the directions in reverse, making sure the metal pin is engaged completely.
Remove the key and put it in a secure place for safekeeping.
Shelves
To prevent damaging the door gasket, make sure to have the door all the way open when pulling the
shelves out of the rail compartment.
The top and bottom shelves can be removed to store larger
items.
When removing a shelf from the rail compartment, make
sure to remove all stored items first. Then move the shelf
to the position where the notch of the shelf is exactly under
the plastic post and lift the shelf. To replace the shelf,
repeat the steps described above in reverse.
The center shelf is fixed in place and cannot be removed. This is
because the glass bead-filled bottle that contains the temperature
sensor for your TMD-02 Thermometer/Alarm is best stored in this
location. The center of your refrigerator or freezer will provide the most
accurate temperature readings.
NOTE: For proper temperature monitoring, keep the bottle in place and do not remove this shelf.