Using the Scope and Meter
Making Automatic Scope Measurements
1
19
Making Automatic Scope Measurements
The test tool offers a wide range of automatic scope
measurements. In addition to the waveforms you can
display four numeric readings: READING 1 4. These
readings are selectable independently, and the
measurements can be done on the input A, input B, input
C or input D waveform.
To choose a frequency measurement for input A, do the
following:
1
SCOPE
Display the SCOPE key labels.
2
F2
Open the READING .. menu.
3
F1
Select the reading number to be
displayed, for example READING 1.
4
ENTER
Select on A. Observe that the
highlight jumps to the present
measurement.
5
ENTER
Select the Hz measurement.
Observe that the top left of the screen displays the Hz
measurement. (See Figure 8.)
To choose also a Peak-Peak measurement for Input B as
second reading, do the following:
1
SCOPE
Display the SCOPE key labels.
2
F2
Open the READING .. menu.
3
F1
Select the reading number to be
displayed, for example READING 2
4
ENTER
Select on B. The highlight jumps
to the measurements field.