Loading ...
Loading ...
Loading ...
Trigger
R&S
®
MXO 4 Series
138User Manual 1335.5337.02 ─ 05
Remote command:
TRIGger:MEVents:SEQuence<se>:COUNt on page 459
Enable reset event
If enabled, the trigger sequence is restarted by the R-trigger condition if the specified
number of B-triggers does not occur before the R-trigger conditions are fulfilled.
Remote command:
TRIGger:MEVents:SEQuence<se>:RESet:EVENt on page 460
Enable reset by time, Reset timeout
If reset timeout is enabled, the instrument waits for the "Reset timeout" time for the
specified number of B-triggers. If no trigger occurs during that time, the sequence is
restarted with the A-trigger.
Remote command:
TRIGger:MEVents:SEQuence<se>:RESet:TIMeout[:ENABle] on page 461
TRIGger:MEVents:SEQuence<se>:RESet:TIMeout:TIME on page 460
7.4 Trigger types
● Edge trigger...........................................................................................................138
● Edge trigger on external trigger source.................................................................139
● Glitch trigger..........................................................................................................141
● Width trigger..........................................................................................................143
● Runt trigger........................................................................................................... 144
● Window trigger...................................................................................................... 146
● Timeout trigger...................................................................................................... 148
● Interval trigger....................................................................................................... 149
● Slew rate trigger....................................................................................................151
● Setup & Hold.........................................................................................................153
● State trigger...........................................................................................................155
● Pattern trigger....................................................................................................... 156
● Line trigger............................................................................................................ 161
7.4.1 Edge trigger
Access: "Menu" > "Trigger" > "Setup" tab > "Type = Edge"
The edge trigger is the most common trigger type. The trigger occurs when the signal
from the trigger source passes the trigger level in the specified direction (slope).
Trigger types