FLUKE 3130-G2M Portable Pneumatic Pressure Calibrator

User Manual for the Fluke 3130-G2M Portable Pneumatic Pressure Calibrator - Page 50

For 3130-G2M.

PDF File Manual, 51 pages, Read Online | Download pdf file

3130-G2M photo
Loading ...
Loading ...
Loading ...
3130
Users Manual
44
Specifications
Specifications are valid for one year unless otherwise noted.
Electrical Specification
Measurement Inputs Range Accuracy
mA 0 mA to 24.000 mA 0.015 % of reading ±0.002 mA
Volts 0 mA to 30.000 V dc 0.015 % of reading ±0.002 V
Outputs Range
mA 4 mA to 24.00 mA
Volts 24 V dc
[1]
[1] Available in mA or V measurement mode
Pressure
Pressure ............................................................ -12 to 300.00 psi
mA .................................................................... 0 to 24.000 mA
Volts ................................................................. 0 to 30.000 V dc
Engineering Units ............................................ psi, bar, mbar, kPa, MPa, kgcm
2
, mmH
2
O
@ 4 °C, mmH
2
O @ 20 °C, cmH
2
O @
4 °C, cmH
2
O @ 20 °C, inH
2
O @ 4 °C,
inH
2
O @ 20 °C, inH
2
O @ 60 °F, mmHg
@ 0 °C, inHg @ 0 °C
Accuracy
Pressure ........................................................... 1 year: 0.025 % of reading ±0.01 % FS
mA .................................................................... ±0.015 % of reading ±0.002 mA
Volts ................................................................ ±0.015 % of reading ±0.002 V
Temperature Effect (all functions) .................. No effect on accuracy on all functions
from 15 °C to 35 °C
Add ±0.002 % F.S. /°C for temps outside
of 15 °C to 35 °C
Loop Power
24 V .................................................................. 23 V to 28 V
Mechanical Specification
Size (H X W X L) ............................................. (387 x 305 x 178) mm, (15.3 x 12.0 x 7.0) in
Weight ............................................................ ~7.0 kg (15.5 lb)
Environmental
Operating Temperature ..................................... -10 °C to +50 °C (14 °F to 122 °F)
Storage ............................................................. -20 °C to +55 °C (-4 °F to +131 °F)
Humidity .......................................................... 10 % to 95 % RH Non-condensing
Altitude ............................................................. 2000 meters, limited by ac adapter