FLUKE 3130-G2M Portable Pneumatic Pressure Calibrator

User Manual for the Fluke 3130-G2M Portable Pneumatic Pressure Calibrator - Page 36

For 3130-G2M.

PDF File Manual, 51 pages, Read Online | Download pdf file

3130-G2M photo
Loading ...
Loading ...
Loading ...
3130
Users Manual
30
Set Up the RS-232 Port for Remote Control
Note
The RS-232 connection cable must be less than 15 m unless the load
capacitance measured at connection points is less than 2500 pF.
Serial parameter values:
1. 9600 baud
2. 8 data bits
3. 1 stop bit
4. no parity
5. Xon/Xoff
6. EOL (End of Line) character or CR (Carriage Return) or both
An RS-232 cable is used for RS-232 communications from the Product to a
computer. If the computer only has USB ports, a USB to RS-232 converter will be
necessary. To connect the Product to a computer, attach the LEMO connector end
of the cable to the pressure module port on the right side of the Product. Then
connect the DB-9 connector to the RS-232 port on the computer. The Product
must be off before you make the connection.
To set up remote operation of the Product with Windows Hyper Terminal, connect the
Product to a COM port on the PC. See Figure 35.
Do the subsequent procedure:
1. Start Hyper Terminal (located on your computer in Accessories/Communications of
the Windows Start menu).
2. Select New Connection.
3. For Name enter “Fluke 3130”.
4. Select the serial port that the Product is connected to.
5. Enter the above information for port settings.
6. Select ASCII setup from File/Properties/Settings and mark these choices:
1. Echo typed characters locally
2. Wrap lines that exceed terminal width
7. Select Ok.
8. To see if the port functions, enter *IDN?. This command shows information on the
Product.