Isopropyl alchohol may be substituted with commonly used, EPA-registered hospital
disinfectant wipes for cleaning purposes if desired. Please refer to the directions
provided by the manufacturer of the wipes for appropriate instructions for use to ensure
effective contact time between the wipe and the device.
It is important to note that solution residue on the magnifying lenses may decrease de-
vice performance. All excess cleaning fluids must be carefully wiped clean if these sur-
faces are cleaned. The lenses of the instrument head may be cleaned with a lint-free
cloth or lens paper.
Steris Sterilization Processes
This product has been validated with the V-PRO 1 Standard Cycle; V-PRO 1 Plus
Lumen and Non Lumen Cycles; V-PRO maX Lumen, Non Lumen, and Flexible
Cycles; V-PRO 60 Lumen and Non Lumen Cycles using the V-PRO 60 Low
Temperature Sterilization System with VAPPROX® HC Sterilant and V-PRO maX Low
Temperature Sterilization System.
Disposable Specula Cleaning/Disinfection Instructions
Disposable specula should not be cleaned or disinfected in any way. Disposable spec-
ula are single patient use only and must be discarded after use. Disposable specula
should not be used if they appear to be visibly contaminated or have accidentally come
into contact with contaminated materials prior to use.
Troubleshooting
The following guidelines should be used to determine if your device has reached its
end of life or requires servicing. If the corrective actions described in this section do
not resolve your issues, please see the warranty section of this manual to have your de-
vice serviced.