ADC 5110NL Otoscope/Ophthalmoscope Diagnostic Set

Use, Care, & Maintenance - Page 4

For 5110NL.

PDF File Manual, 13 pages, Read Online | Download pdf file

5110NL photo
Loading ...
Loading ...
Loading ...
4
CAUTION: Do NOT immerse the battery handle in liquid. Consult the cleaning and dis-
infection instructions for each part for further information about cleaning and disinfecting
this product. Never clean or disinfect any device with batteries in it.
CAUTION: Federal law restricts this device to sale by or on the order of a physician or li-
censed healthcare practitioner.
CAUTION: Turn off instrument when not in use. Avoid powering on without an instrument head
attached. When storing this device for extended periods of time, remove the batteries.
WARNING: The carrying cases provided with these devices are intended for long term
storage or transport between facilities or when shipping devices to and from ADC’s facility
for repair or servicing. Carry cases (including internal liners) can NOT be cleaned or dis-
infected and should be safely discarded if contaminated. Do not carry the case into a con-
taminated environment.
WARNING: When replacing lamp, allow lamp to cool for five minutes before handling.
WARNING: Only the approved sterilization method included in this IFU has been vali-
dated for this device or its corresponding components. Any other sterilization method
may compromise the safety and effectiveness of this device. (This includes steam steril-
ization.)
WARNING: Do NOT place devices or accessories back into carrying case after use on a
patient without first cleaning/disinfecting the device/accessories as described in the clean-
ing and disinfection section of this manual. Carry case liners can NOT be cleaned or dis-
infected and should be discarded if contaminated.
WARNING: Use of any accessories or materials not indicated in the user's manual can
degrade the minimum safety of the equipment.
WARNING: Only replace listed accessories (attachments), lamps, or batteries for the device.
WARNING: During use metal components near instrument head can become warm. This
is especially true if device is on for extended periods of time. Do not leave device on when
not in use.
WARNING:The light emitted from this instrument is potentially hazardous. The longer the
duration of exposure, the greater the risk of ocular damage. Exposure to light from this
instrument, when operated at maximum intensity, will exceed the safety guideline after
3 minutes and 47 seconds. Exposure times are cumulative for a 24 hour period.
Loading ...
Loading ...
Loading ...