HANTEK6004BD 6004BD Series

Hantek6000BD Manual - Page 18

For HANTEK6004BD.

PDF File Manual, 82 pages, Read Online | Download pdf file

HANTEK6004BD photo
Loading ...
Loading ...
Loading ...
Hantek 6000BC/Hantek 6000BD/Hantek 6000BE
Digital Oscilloscope
USER’S MANUAL 18
3. Attach the tip of probe and ground nip to the Connector of Probe compensator. Click the
button. A square wave will be displayed within a several seconds. (Approximately 1
kHz, 2V, peak- to- peak).
4. Inspect CH2 with the same method. Repeat steps 2 and 3.
1.7 Self Calibration
The self calibration routine lets you optimize the oscilloscope signal path for maximum
measurement accuracy. You can run the routine at any time but you should always run
the routine if the ambient temperature changes by 5v or more. For accurate calibration,
power on the oscilloscope and wait twenty minutes to ensure it is warmed up. To
compensate the signal path, disconnect any probes or cables from the input connectors.
Then, access the Utility -> Calibration” option and follow the directions on the screen.
The self calibration routine takes about several minutes.
Loading ...
Loading ...
Loading ...