Hanna HI5222 Laboratory Research Grade Two Channel Benchtop Ph/mv/ise Meter

User Manual - Page 57

For HI5222.

PDF File Manual, 84 pages, Read Online | Download pdf file

HI5222 photo
Loading ...
Loading ...
Loading ...
57
To return to normal Measure mode press
.
Note: If the reading is out of range, “----” will be displayed on the LCD.
KNOWN ADDITION
To measure the concentration of a sample using the
Known Addition incremental method:
• Press
and then to select ISE
Measure mode for the selected channel.
Select the Known Addition method (see ISE Setup for
details).
Prior to starting a KA procedure, the ISE sensor must
be calibrated with a minimum of two standards
containing ISA. The slope of the electrode will be
used in all calculations involved in KA.
If following an established procedure: Press
,
Edit the method variables and follow the procedure
below.
• Press
to set the method parameters. Press
/ to select next/previous parameter
to edit, then press and use or
to set the desired parameter value. Press
to save the modified value and then press
to exit method parameters edit menu.
If developing a procedure: Before attempting Known
Addition analysis it is important to determine what
sample volume, standard concentration and standard
volume will produce the best results. As a general rule,
the addition of standard should change the mV value
of the sample by 15 - 20 mV. For a positively charged ion (i.e. Sodium, Potassium, Calcium), the
standard addition should increase the mV. For a negatively charged ion (i.e. Sulfide, Fluoride,
Chloride), the standard addition should decrease the mV. Start with a small trial. For example:
Measure 50 mL of sample, add a magnetic stir bar, place on a stirrer, add ISA (consult ISE
manual) and place ISE electrode tip into the sample. Put instrument in mV mode and record
the observed mV. Using a micropipette, add a volume of the highest ISE standard available
(i.e. 0.1M or 1000 ppm).
ISE MEASUREMENT (HI5222 only)
Loading ...
Loading ...
Loading ...