42
• Use
or to select the Calibration option.
• Press
and use or to highlight
the Edit Standard Group option.
Press
and use and to choose
the ISE standard to be included in the standard group.
• Press
/ to add/remove the selected ISE
standard to/from the standard group.
• Press
to return to Calibration options and to
save the changes.
Calibration Reminder - See Calibration option from pH
Setup section.
Set Reminder Period - See Calibration option from pH
Setup section.
Clear Calibration - See Calibration option from pH Setup section.
Electrode Type
This option allows the user to select the desired Ion Selective Electrode used for measurements from
a list: Ammonia, Bromide, Cadmium, Calcium, Carbon Dioxide, Chloride, Cupric, Cyanide, Fluoride,
Iodide, Lead, Nitrate, Potassium, Silver, Sodium, Sulfate, Sulfide and five custom ISE. For the standard
ISE it is possible to view the Ion constants (Name, Molar Weight and Electric Charge/Slope), while for
the custom ISE all these constants can be manually set.
To set the Electrode Type:
Press
while in ISE Measure mode.
Press
.
Use
or to select the Electrode Type
option.
• Press
and use or to select
the desired standard ISE or a custom one from the list.
For standard ISE:
• Press
to visualize the Ion Constants and then
press
at any time to exit Ion Constants view
mode.
ISE SETUP (HI5222 only)