4.3 Long press to enter calibration mode; Short press to exit.
4.4 Insert the probe in the 7.00 pH calibration solution; Stir gently;
leave it to stand; Wait for the smiley face to appear and
stay on the screen (see Diagram 3); short press to complete
1
st
point calibration, tester returns to measurement mode;
calibration icon displays on button left side of the screen.
4.5 Rinse probe in distilled water. Long press to enter calibration
mode; insert the probe in the pH 4.00 calibration solution, stir gently; leave it
to stand; wait for the smiley face to appear and stay on the screen; then
short press to complete the 2
nd
point calibration,
tester returns to measuring mode, calibration icons display on
bottom left side of LCD.
Notes
a) Tester will automatically recognize pH buffer solution, users can choose
calibration points: 1 point, 2 points, or 3 points. But the 1st point calibration
must be in calibration solution of 7.00 pH, then followed by 2nd or 3rd point
calibration. For details, please refer to the following table:
Calibration Solution
Calibration
Indication icon
Recommended Accuracy
and Range
1-point
Calibration
7.00 pH
Accuracy 0.1pH
2-point
Calibration
7.00 pH and 4.00 pH
Measuring Range <7.0
pH
7.00 pH and 10.01 pH
Measuring Range>7.0
pH
3-point
Calibration
7.00 pH, 4.00 pH
and 10.01 pH
Wide measuring range
b) Automatic self-diagnostic information: if measured value is far off to the
preset range, LCD will displayEr1; In calibration mode, if the measured
Diagram - 3