ITECH IT6123B 72V 1.2A 86.4W Programmable DC Power Supply

Manual - Page 31

For IT6123B.

PDF File Manual, 47 pages, Read Online | Download pdf file

IT6123B photo
Loading ...
Loading ...
Loading ...
Basic operation
Copyright © ITECH Electronic Co., Ltd. 24
When the value in operation register changes, the TTL level of Out + and Out
will change simultaneously.
The default set of config menu
Memory
group 0
Out_Recal
Off
Set_Recall
On
Buzzer
Off
Knob
Off
Comm
RS232
Port
Trigger
Trig
Mauual
RI
Off
DFI
Off
Limit voltage (Limit_Volt)
Limit voltage range is 0V - Vmax.
Limit voltage is used to limit the voltage setting for a certain device under test.
For example: the limit voltage is set to 12V, when you set voltage to 15V (over 12V),
then the voltage will auto adjust to 12V.
The operation to set voltage limit:
1. Press (Shift) +
I-setI-set
(Menu) to menu
2. Press to System menu, press
Enter
to confirm
3. Select Limit_Volt in System menu, Press
Enter
to set the voltage limit
The default set of Limit_volt is the max rated voltage.
ON TIMER STATE
This item is used to set the “time on- load” function. In ON mode, the indicator light
“Timer” will be lit on the VFD screen. When output of power supply is enabled,
timer will begin to work, after reaching the definite time, output will be off
automatically. If in OFF mode, the timer function will not be enabled. The default
set is OFF.
RESET
This item is used to reset all items in the menu. If you select >YES, then unit will
restored to factory setting. If you select >NO, all setting in the menu will remain
unchanged.
OutMode (IT6162B/IT6164B Specific function)
This item is used to set the power supply output mode, including Volt-Wave Prio
(voltage priority) mode and Curr-Wave Prio (current priority) mode. Under voltage
mode, when output wave quality is maintained, current may overshoot; under
current mode, when current output wave quality is maintained, voltage rise speed
will get slower.
Loading ...
Loading ...
Loading ...