AOPUTTRIVER 8542039922 Digital Clamp Meter AP-570N, 6000 Counts Auto-ranging Multimeter

User Manual

For 8542039922.

PDF File Manual, 2 pages, Read Online | Download pdf file

8542039922 photo
6000 DIGITS DISPLAY
AC/DC AUTO CAL CLAMP MULTIMETER
OPERATION MANUAL
This LCD Auto Range & Auto Power off Digital AC/DC AUTO
CAL clamp multimeter is a portable, 6000 digit multimeter. It is
ideally suited for field, laboratory, shop and home applications.
1. SAFETY INFORMATION
The following safety information must be observed to insure
maximum personal safety during the operation at this meter.
1) When measuring voltage ensure that instrument is not
switched to the current range, resistance range, diode and
continuity range, capacitance range or temperature range.
2) Use extreme care when measuring voltage above 50V.
especially from sources where high energy is existed.
3) Avoid making connections to "live" circuits whenever possible.
4) Before making resistance measurements, diode or continuity
test, capacitance test or temperature test, ensure that the
circuit under test is de-energized.
5) Always ensure that the correct function and range is selected.
6) Extreme care should be taken when using the instrument to
conjunction with a current transformer connected to the
terminals if an open circuit occurs.
7) Ensure that the test leads and probes are in good condition
with no damage to the insulation.
8) Take care not to exceed the over-load limits as given in the
specifications.
9) Before opening the cover of the battery cabinet to replace
batteries. disconnect the test leads from any external circuit,
set the selector switch to "OFF" position.
10) Keep the fingers after the protection ring when measuring
through the instrument lead.
11) Change the battery when the symbol appears to avoid
incorrect data.
2. Panel Layout
1) Rotary Switch: use this switch to select functions and ranges.
2) D.HOLD/ Back Light keyIn any range, push the key, the
present display value will be locked and the " H " symbol will
appear, push it again to exit HOLD and the " H"symbol
disappear. Press “D.HOLD” button more than 2 seconds, the
back light will light, press it more than 2 seconds again, the
back light will light off or After 15 seconds Auto light out.
3) SELECT key:This key work on the "CAP Ω" range , Push
the key to choose resistance, diode ,continuity or capacitance
test, on the voltag or current range,change to DC/AC, and /
range , change to /
4) MAX/MIN keyPush the key to select MAX mode, push it
again to change MIN mode, push once again to change
max-min,press the key for more than 2 seconds to go back
auto range mode. And Push the key to change manual range
mode .But in Hz/Duty and Capacitance measurement, it can
not use.
5) Hz/DUTY keyIn ACV/ACA or Hz range, push the
key, you can measure the Hz ,push again, can measure the
duty.
6REL keyPush the key, the present display value will be stored
in memory, then the new display value is the difference
between input value and stored data. In Hz/Duty measurement,
it can not work. press the key for more than 2 seconds to
measure inrush current at ACA range
7) LCD DisplayLCD Dual Display, facilitates reads the data.
8)T+V Ω Input Jack COM Input JackT- Input Jack.
2. SPECIFICATIONS
2.1 GENERAL SPECIFICATIONS
Display: 6000 digit LCD with a max. reading of 6000.
Range control: Auto range control.
Polarity: Automatic negative polarity indication.
Zero adjustment: Automalic.
Overrange indication: Only the "OL" display.
Low battery: The " " is display when the battery voltage is below 6.2V.
Auto Power Off: 15 minutes after stopping the switch or no push
button, the meter automatically enter to power off mode.
Press the "SELECT" button to turn on and cancel the
automatic shutdown.
Safety Standards: The meter is up to the standards of IEC1010
Double Insulation, Pollution Degree 2, Overvoltage
Category III 1000V.
Clamp opening size: 45mm.
Operating Environment: Temperature 32104°F(040),
humidity80%RH.
Storage Environment: Temperature -4140°F(-2060),
humidity90%RH.
Power supply: 9V Zinc-carbon battery.
Dimension: 225(H)×97(W40(D)mm.
Weight: Approx. 350g (including batteries).
2.2 ELECTRICAL SPECIFICATIONS
Accuracies are ±(% of reading +numder in last digit) at 23±5,
70%RH.
2.2.1 DC Voltage
Range
Accuracy
Resolution
600mV
1.0% of rdg+5digits
0.1mV
6V
1mV
60V
10mV
600V
100mV
1000V
1.5% of rdg+5 digits
1V
Overload protection: 1000V DC/750Vrms AC
Impedance: 10MΩ,More than 100MΩ on 400mV scale
2.2.2 AC Voltage(RMS)
Range
Accuracy
Resolution
Frequency
600mV
3.0% of rdg+15 digits
0.1mV
40100Hz
6V
1.5% of rdg+5 digits
1mV
40400Hz
60V
10mV
600V
100mV
750V
2.5% of rdg+5 digits
1V
40100Hz
Average sensing, calibrated to rms of sine wave
Overload protection: 1000V DC/750Vrms AC
Impedance: 10MΩ.
2.2.3 DC Current
Range
Resolution
Frequency
600A
0.1A
5060Hz
1000A
1A
Overload protection: 1000Arms within 60 seconds
2.2.4 AC Current
Range
Accuracy
Resolution
Frequency
600A
3.0% of rdg+10 digits
0.1A
5060Hz
1000A
3.5% of rdg+10 digits
1A
Average sensing, calibrated to rms of sine wave
Overload protection: 1000Arms within 60 seconds
2.2.5 Resistance
Range
Accuracy
Resolution
600Ω
1.0% of rdg+8 digits
0.1Ω
6kΩ
60kΩ
10Ω
400kΩ
100Ω
6MΩ
1kΩ
60MΩ
2.5% of rdg+15 digits
10kΩ
Overload protection: 250V DC/250Vrms AC
2.2.6 Capacitance
Range
Accuracy
Resolution
9.999nF
2.5% of rdg+15 digits
1PF
99.99nF
10PF
999.9nF
2.5% of rdg+10 digits
100PF
9.999μF
1nF
99.99μF
10nF
999.9μF
3.0% of rdg+20 digits
100nF
9.999mF
5.0% of rdg+25digits
1uF
99.99mF
10uF
Overload protection: 250V DC/250Vrms AC