Loading ...
Loading ...
Loading ...
11
9. Cleaning and storing the unit
Clean your blood pressure device carefully with a slightly
damp cloth only.
Do not use any cleaning agents or solvents.
Under no circumstances should you hold the unit under
water, as this can cause liquid to enter and damage the
unit.
If you store the unit, no heavy objects should be placed
on top of it. Remove the batteries. The cuff line should not
be bent sharply.
10. Technical specifications
Model no. SBM 42
Measurement
method
Oscillometric, non-invasive blood
pressure measurement on the upper
arm
Measurement range Cuff pressure 0-300 mmHg,
systolic 60-260 mmHg,
diastolic 40-199 mmHg,
Pulse 40-180 beats/minute
Display accuracy Systolic ± 3 mmHg,
diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement inac-
curacy
Max. permissible standard deviation
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
Memory 4 x 30 memory spaces
Dimensions L 134 mm x W 102 mm x H 52,5 mm
Weight Approximately 280g (without batter-
ies and cuff)
Cuff size 22 to 36 cm
Permissible operating
conditions
+10 °C to +40 °C, ≤ 85 % relative air
humidity (non-condensing)
Permissible storage
conditions
-20 °C to +50 °C, ≤ 85 % relative air
humidity, 800-1050 hPa ambient
pressure
Power supply 4 x 1,5 V
AA batteries
Battery life For approx. 500 measurements,
depending on the blood pressure
level and/or pump pressure
Accessories Instruction for use, 4 x 1.5 V AA bat-
teries, storage pouch
Classification Internal supply, IPX0, no AP or
APG, continuous operation, type BF
applied part
Downloaded from www.ManualsFile.com manuals search engine
Loading ...
Loading ...
Loading ...