AccuCold, Div of Felix Storch, Inc. | R005-07102018
4
Buttons Description
Default Data Logger Settings
Code
Function
Range
Default Setting
*Please enter the correct temperature units
o
F /
o
C
C1
High temp. alarm
C2 ~ 100
o
C /212
o
F
8.0
o
C
C2
Low temp. alarm
-45
o
C /-49
o
F ~ C1
2.0
o
C
C3
Alarm hysteresis
0.120.0
o
C
0.236
o
F
1.0
o
C /2.0
o
F
C4
Alarm delay
0090 min
0 min
C5
Start delay
0090 min
0 min
CF
Temperature Unit
o
C =Celsius
o
F =Fahrenheit
o
C
E5
Offset temperature
-2020
o
C
-3636
o
F
0.0
o
C/
o
F
L1
Logging interval
00240 min
05 min
PAS
Password
00 ~99
50
Programming the Data Logger
Password Input
From the main display screen:
Hold the SET key for 3 seconds. Use the up and down arrows to adjust the
password to the correct password. By default the correct password is 50.
Press SET key ONCE to confirm the settings.
High Alarm
Temperature Setting
By default, the high and low alarm settings are 8
o
C and 2
o
C respectively. To reset high
alarm and low alarm temperature settings, follow instructions below.
From the main display screen:
Hold the SET key for 3 seconds. Enter the correct password then press the SET key
ONCE to enter the HI Temp Alarm setting mode. Use the up and down arrows to
adjust the temperature accordingly. Press SET key ONCE to confirm the settings.
REC/STOP
Press REC/STOP to STOP or RECORD data.
MAX/MIN
Press for 3 seconds to ERASE the MIN and MAX temperature
history.
DL
Copy the recorded data (CSV file) to USB
SET
Hold the SET button to cycle through configuration settings.
Up/Down keys to change settings. Press and hold either key to
advance the values rapidly.