For ARG8PV-CRTTEST. Also, The document are for others Summit models: ARS1PV-CRT, ARG1PV-CRT, ARS3PV-CRT, ARG3PV-CRT, ARG31PVBIADA-CRT, ARS32PVBIADA-CRT, ARS6PV-CRT, ARG6PV-CRT, ARG61PVBIADA-CRT, ARS62PVBIADA-CRT, ARS8PV-CRT, ARG8PV-CRT, ARS12PV-CRT, ARG12PV-CRT, ARS15PV-CRT, ARG15PV-CRT, ARS18PV-CRT, ARG18PV-CRT

PDF File Manual, 44 pages, Read Online | Download pdf file

ARG8PV-CRTTEST photo
Loading ...
Loading ...
Loading ...
13
F0
Defrost Type
00=defrost by
turning of comp.
00
C11
Comp.
Force
Running
Time
00=comp. Stop /
01-99min=starting
time
0 min
F1
Max defrost
duration
01~60 min.
20 min.
C12
Alarm
Output Type
00=contact
actuation when
alarm / 01=contact
disconnect when
alarm
01
F2
Defrost
Interval
XXhr/time
00~24 hr.
00
C13
Normal
State temp.
display type
00=liquid bottle
temp / 01=cabinet
air temp.
00
F4
Temp.
displaying on
screen when
defrost
00= normal Temp;
01=latest Temp.
02=Fixed Temp. ;
03=showing “dEF”
on screen
00
CF
Temperature
Unit
°F=Fahrenheit
°C=Celsius
°F
C1
High Temp
Air Alarm
78.0°F
26.0°C
77.0°F / 25.0°C
do1
00=do not alarm /
01=99
min=delayed alarm
02 min
C2
Low Temp Air
Alarm
67.0°F
19.0°C
68.0°F / 20.0°C
do2
Comp.
status when
door open
0=stop /
01=original status
0 1
do3
Light status
when door
open
00=start /
01=original status
0 1