Summit ARG8PV-CRTTEST 8 Cu.Ft. Upright Controlled Room Temperature Cabinet

User Manual - Page 12

For ARG8PV-CRTTEST. Also, The document are for others Summit models: ARS1PV-CRT, ARG1PV-CRT, ARS3PV-CRT, ARG3PV-CRT, ARG31PVBIADA-CRT, ARS32PVBIADA-CRT, ARS6PV-CRT, ARG6PV-CRT, ARG61PVBIADA-CRT, ARS62PVBIADA-CRT, ARS8PV-CRT, ARG8PV-CRT, ARS12PV-CRT, ARG12PV-CRT, ARS15PV-CRT, ARG15PV-CRT, ARS18PV-CRT, ARG18PV-CRT

PDF File Manual, 44 pages, Read Online | Download pdf file

ARG8PV-CRTTEST photo
Loading ...
Loading ...
Loading ...
12
Table of Parameters and Factory Settings
Parameter
Function
Set Range
Default
Parameter
Function
Set Range
Default
Set
Setpoint
72.0°F
22.5°C
72.0°F
22.5°C
C3H
Product
Temp. High
Temp. Alarm
77.0°F
25.0°C
77.0°F
25.0°°C
PAS
Password
00~99
15
C3L
Product
Temp. Low
Temp. Alarm
68.0°F
20.0°C
68.0°F
20.0°C
E1
Lowest set
68.0°F
68.0°F / 20.0°C
C4
Hysteresis
0.1~20.0°C /
0.1°C /
point limit
20.0°C
Alarm
1.8~36.0°F
0.18°F
E2
Higher set
point limit
77.0°F
25.0°C
77.0°F / 25.0°C
C5
Start-up
Temp. Alarm
delay
00~99 min
60 min
E3
Temp.
Hysteresis
0.1~20.0°C /
0.2~36.0°F
2.0°C / 3.6°F
C6
Temp. Alarm
Delay
00~99 min
0 min
E40
Turning On
delay time
00 ~10 min.
3 min.
C7
Power Off
relay Alarm
00=do not alarm /
01= alarm
01
E41
Comp. start
delay time
00 ~10 min.
3 min.
C8
Alarm relay
close after,
muffle alarm
relay switch
00=open /
01=close
00
E5
Offset on Air
Temperature
1.8°F/ 1.0°C
1.8°F/ 1.0°C
C9
Restart time
after buzzer
mute
00=do not start /
01 ~30 min=restart
time
10 min
E6
Offset on
Product
Temp.
-10.0~10.0°C / -
18.0~18.0°F
0.0°C / 32°F
C10
Comp.
Force stop
time
01~99 min
99 min