2.2.4
ACA
RANGE ACCURACY RESOLUTION
2mA 1uA
20mA
±(1.0%+5)
10uA
200mA ±(2.0%+5) 100uA
20A ±(3.0%+10) 10mA
Max measuring voltage drop200mV.
Max Input current: 20A (less than 10 seconds).
Overload protection0.2A/250V fast-blow fuse, 20A/250V
quick-action fuse for range 20A.
Frequency response: 40Hz-200Hz.
Display: Sine wave RMS (AVG value response).
2.2.5 RESISTANCE
RANGE ACCURACY RESOLUTION
200 ±(0.8%+5) 0.1
2K 1
20k 10
200k 100
2M
±(0.8%+3)
1K
20M ±(1.0%+15) 10k
2000M
±(5.0%(reading-
10)+20)
1M
Open voltage: less than 3V.
Overload protection: 250V DC or AC peak value.
NOTE:
1. At 200 range, the test leads should be short-circuited, and
measure the lead to lead resistance, then, subtract from the
real measurement.
2. It is normal to display 10M when the test leads shorted in
range 2000M, it will not effect the accuracy and shall be
subtracted from the value measured. For example: The object
resistance is 1000M, the reading value is 1010M, then the
correct value shall be 1010-10=1000M
3. Lagged display of value is normal when measuring
resistance higher than 1M. Please wait until the display is
stable.
3