SIGLENT
102 SDG1000X User Manual
2.13.4 Output
Press Utility Output to enter the following interface.
Figure 2-60 Output Setup Interface
Load
For the [CH1] and [CH2] connectors on the front panel, the generator has an
output impedance of 50Ω. If the actual load does not match the source
impedance, the displayed voltage will not be the same as the output voltage
setting on the generator. This function is used to match the displayed voltage
with the expected one. This setting does not actually change the output
impedance to any other value.
Steps for setting the load:
Press Utility Output Setup Load, to set the output load. The load
parameter shown on the down bottom is the default setting when the power is
on or the pre-set load value.
High Impedance: displayed as HiZ;
Load: the default is 50Ω and the range is 50Ω to 100kΩ.