Loading ...
Loading ...
Loading ...
![](https://files.manualsfile.com/3336055-sna5012a/bg4e.png)
SNA5000A Vector Network Analyzer User Manual 76
Figure 3-26 Set the interface
Power calibration operation steps:
1. Connect the port-under-test to a compatible power meter (Keysight U2000 series, R&S
NRP series, etc..) and probe to ensure normal communication between the vector
network analyzer and the power meter.
2. Preset vector network analyzer.
3. Set internal source output power, IF bandwidth, scan points, test frequency band and other
parameters.
4. Select Cal→Power Cal to enter the Settings screen.
5. Select the test channel and select the power output port.
6. Set parameters such as power offset, capacity limit, reading times, maximum iteration
times, etc.
7. Click Take Cal Sweep to start the power output measurement.
8. Click OK to complete the calibration and exit the power calibration interface after the
measurement is completed.
Loading ...
Loading ...
Loading ...