31 SNA5000A Vector Network Analyzer User Manual
Operating steps:
Press Sweep, use the knob or arrow keys to focus on the Sweep→Number of Points
parameter item, Enter the number of points required, Press ENTER to exit editing mode.
The number of data points collected by the vector network analyzer during the
measurement sweep can be set to any number between 1 and 100001.
Note: Maximum point limits may differ for some measurement classes.
• For maximum trace resolution, use the maximum data points.
• For faster throughput, use a minimum number of data points to provide an acceptable
resolution.
• To get the best number of points, look for values that do not differ significantly in the
measurement as you add points.
• To ensure accurate measurement calibration, ensure that the user uses the same number
of points for calibration and measurement.
• Points are the number of data items collected in one sweep. This number can be set
separately for each channel.
• To obtain high trace resolution for excitation values, select a larger point value.
• For high throughput, keep the smaller point value within the allowable trace resolution
range.
• For high measurement accuracy after calibration, use the same number of points as the
actual measurement.
2.4.2 Sweep type
Key operation:
Press Sweep, use the knob or arrow keys to focus on the Sweep→Sweep Type parameter
item. Press ENTER or Knob to enter the editing state, then move the cursor to the specified
position by up and down direction keys or Knob, set the sweep type to the required
measurement type, press ENTER or Knob to select the current option.
Sweep type:
• Linear frequency sweeping
• Log frequency sweeping
• Power sweeping
• CW time sweep
• Segmented sweeping
Linear frequency sweep: Set the measured abscissa frequency scale as a linear scale, and
keep the full frequency scale at equal intervals.