40
Error
message
Possible
cause
Solution
The cu was
not attached
correctly.
Please observe the notes
in chapter "7. Usage"
under the section "Atta-
ching the cu".
During a new measure-
ment, check whether the
cu can now be inflated
properly.
An error oc-
curred during
measurement.
Please wait one minute
and repeat the measu-
rement.
Ensure that you do not
speak or move during
the measurement.
The inflation
pressure is
higher than
300 mmHg or
the measured
blood pressure
is outside the
measuring
range.
The batteries
are almost
empty.
Insert new batteries into
the device.
Error
message
Possible
cause
Solution
The data could
not be sent via
Bluetooth
®
.
Proceed as described in
chapter „7. Usage“ under
the section
„Transfer via Blue-
tooth
®
“ into the manual
memory mode and try to
restart the data transfer.
A device error
has occurred.
Please repeat the mea-
surement after a pause
of one minute.
If the problems persist despite the proposed corrective
actions, please contact customer service.
10. Disposal
Repairing and disposing of the device
• Do not repair or adjust the device yourself. Proper op-
eration can no longer be guaranteed in this case.
• Do not open the device. Failure to comply will result in
voiding of the warranty.
• Repairs must only be carried out by Customer Services
or authorised suppliers. Before making a claim, please
check the batteries first and replace them if necessary.