10
Programming the Alarms
The thermometer displays the internal temperature of the unit being monitored, as well as the
maximum and minimum temperatures reached. The maximum and minimum readings reflect the
highest and lowest temperatures since the battery was inserted or since the MODE/RESET key was
last pressed. If the temperature rises above or falls below the set temperature range, the alarm will
sound. By default, the high alarm temperature setting is 8°C (46.4°F) and the low alarm temperature
setting is 2°C (35.6°F). To reset these limits, follow these instructions:
NOTE: Set high and low alarm temperature limits above and below the refrigerator/ freezer
operating temperature to prevent false alarms.
High Alarm Temperature Setting: Press the HIGH/LOW key ONCE, then press MODE/RESET
to enter the High Alarm temperature setting mode. Use the Up and Down arrows to adjust the
temperature accordingly. Press MODE/RESET once again to confirm the setting.
Low Alarm Temperature Setting: Press the HIGH/LOW key TWICE, then press MODE/RESET
to enter the Low Alarm temperature setting mode. Use the Up and Down arrows to adjust the
temperature accordingly. Press MODE/RESET ONCE to confirm the setting.
By default, the low and high alarms should be audible. To turn alarms on/off:
Audible High Alarm Setting: To make sure the high alarm is audible, press HIGH/LOW key
ONCE (to activate HI-SET on the screen). Then press the ALERT ON/OFF key ONCE to
activate the HIGH-ALARM beeping sound. Press again to deactivate.
Audible Low Alarm Setting: To make sure that the low temperature alarm is audible, press the
HIGH/LOW key TWICE to activate LOW-SET on the screen. Then press the ALERT ON/OFF
key ONCE to activate the LO-ALARM beeping sound. Press again to deactivate.
NOTE: For an activated high alarm and low alarm buzzer, the display should indicate HI-ALARM
and LO-ALARM (see illustration on previous page).
AGP96RF and AGP34RF Series
The AGP96RF and AGP34RF series ships with a non-calibrated
alarm/thermometer. For models with suffix “LCAL,” the device is
calibrated in our ISO/IEC 17025:2017 laboratory and will include
a 2-year certificate. Both alarms can be attached to the front
door. The temperature probe is incased in a glycol bottle and
should be placed inside the refrigerator section, ideally towards
the center of the unit. The door of the refrigerator or freezer can
be closed over the probe cord.
The alarm instructions will be packed with the alarm. For technical support, please call 800-932-