10
The green movement indicator light shows that the control unit is turned
on and monitoring baby (it does not represent babies breathing rate).
Acoustic and visual red alarm light are activated if baby's breathing
movement rate stops or slows significantly (below 10 breaths per
minute).
If red low-battery warning light flashes, replace all batteries as soon as
possible. DO NOT use rechargeable batteries.
Turn off by pressing the On/Off button before taking your baby out of cot.
Make sure the mattress is in contact with the sensor pads.
NOTE: A polystyrene foam mattress could cause false alarms.
Daily Testing Operation
You must conduct this quick performance test daily.
The performance test should be conducted while baby is not in
cot and if there are any fans, air conditioning units, cot mobile,
humidier, ceiling projector or other devices in the babies
room, please ensure you conduct the daily performance test
with these devices turned on if applicable.
Turn device on and leave the room.
• Ensure that alarm goes off within 20 seconds (due to absence of
breathing movements).
• In the case that the alarm does not sound within 20 seconds refer to
troubleshooting on page 14.
The alarm may sound with or without preliminary ticking.
• Check that alarm can be heard in all locations throughout the home or
place of care.
This performance test must be conducted daily and every time you
move the monitor, change bedding or relocate basenet/cot position.
IMPORTANT: You should not rely on Babysense7 in any location
throughout the home or place of care where the alarm cannot be
heard.
Babysense7 is an exceedingly sensitive device which monitors baby's
breathing movements through the mattress during sleep. There are two
main causes of malfunction.
Operation
Loading ...
Loading ...
Loading ...