Table 2
GUIDANCE AND MANUFACTURER’S DECLARATION –
ELECTROMAGNETIC IMMUNITY
IMMUNITY TEST
IEC 60601-1-2
TEST LEVEL
COMPLIANCE LEVEL
Electrostatic discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±2 kV, ±4kV, ±8 kV, ±15
kV air
±8 kV contact
±2 kV, ±4kV, ±8 kV,
±15 kV air
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV signal
input/output
100 kHz repetition
frequency
±2 kV for power supply
lines
±1 kV signal
input/output
100 kHz repetition
frequency
Surge
IEC61000-4-5
±0.5 kV, ±1 kV
differential mode
±0.5 kV, ±1 kV, ±2 kV
common mode
±0.5 kV, ±1 kV
differential mode
±0.5 kV, ±1 kV, ±2 kV
common mode
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
0% U
T
; 0,5 cycle. At 0°,
45°, 90°, 135°, 180°,
225°, 270° and 315°.0%
U
T
;
1 cycle and 70% U
T
;
25/30 cycles;
Single phase: at 0°.0%
UT; 250/300 cycle
0 % U
T
; 0,5 cycle. At 0°,
45°, 90°, 135°, 180°,
225°, 270° and 315°.0 %
U
T
;
1 cycle and 70% UT;
25/30 cycles;
Single phase: at 0°.0 %
U
T
; 250/300 cycle
Power frequency
magnetic field
IEC 61000-4-8
30 A/m
50 Hz/60 Hz
30 A/m
50 Hz/60 Hz