Loading ...
Loading ...
Loading ...
11
FEATURES AND FUNCTIONS
Functions setting
(1) Inflation level setting
AUTO (Automatic setting): Can be used when the SINGLE, AVG, or MANU Mode is selected.
The unit estimates the systolic blood pressure during inflation and inflates to a proper cuff pressure
(approximately 30-40 mmHg above the patient’s systolic pressure).
Manual level setting: Inflation level can be set manually between 100 and 280 mmHg. Set the
level to 30 to 40 mmHg higher than the expected systolic pressure.
• To set the P-SET to “AUTO”, turn the P-SET Knob counterclockwise as far as it goes until you can hear a click.
• In the “AUTO” setting, inflation level may not be set automatically when the systolic blood pressure is more than
220 mmHg. Use the unit in the manual setting.
• If the cuff has not been inflated to the necessary level, it may be re-inflated automatically.
ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
k
P
-
S
ET
100
AUTO
1
40 180
220
260
280
M
ODE
AVG .
MANU.
CHECK
SINGLE
ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
k
P
-
SET
100
AUTO
140 180
220
260
280
MODE
AVG .
MANU.
CHECK
SINGLE
(2) Non-display function
Use to prohibit the display of measurement results. However,
the cuff pressure during measurement is displayed.
This function can be used in the SINGLE and AVG Modes.
By pushing the Button, display or non-display of status
is switched alternatingly.
H
IDE
(3) Manual deflation control
Accelerate deflation by pushing the DEFLATION (deflation
control) / Measurement Result Display Switch Button during
deflation in the measurement by the MANU Mode.
With each push of the Button, cuff is deflated rapidly
in increments of 5 to 10 mmHg.
BPM
mmHg
PULSE
SYS
mmHg
DIA
AC
AC
Display Non-display
HIDE
HEM-907-Z2_H_M_HEM-907-Z2_H_M.qxd 31/01/2013 15:59 Page 11
Loading ...
Loading ...
Loading ...