Instructions for Use for KardiaMobile 6L
Page 16 of 18
Guidance and manufacturer’s declaration—electromagnetic immunity
KardiaMobile 6L is intended for use in the electromagnetic environment specified below. The customer
or the user of KardiaMobile 6L should assure that it is used in such an environment.
Immunity
test
IEC 60601
test level
Complian
ce level
Electromagnetic environment - guidance
Radiated RF
IEC 61000-4-
3
10 V/m
80 MHz to
2.7 GHz
10 V/m
Portable and mobile RF communications
equipment should be used no closer to any part
of KardiaMobile 6L, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
< 80MHz
80 MHz to 800 MHz
800 MHz to 2.7 GHz
where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in
each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people
a
Field strength from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which KardiaMobile 6L is used exceeds the applicable RF compliance level
above, KardiaMobile 6L should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating KardiaMobile 6L.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.