* Note: If your systolic or diastolic pressure is outside the standard range (above 135/85 mmHg) the Heartbeat symbol ( ) will blink. Please refer to Section 3.3.

Notes:
) appears on the display, turn the monitor off then replace all batteries at the same time. Disposal of used batteries should be carried out in accordance with the national regulations for the disposal of batteries.
Remove tight-fitting clothing or tight rolled up sleeve from your upper arm. Do not place the arm cuff over thick clothes.
1. Insert the air plug into the air jack securely

2. Put your arm through the cuff loop.

The bottom edge of the arm cuff should be 1 to 2 cm above the elbow. Marker (arrow under the air tube) is centred on the middle of your inner arm.

3. Close the fabric fastener FIRMLY.

Notes:


To take a measurement, you need to be relaxed and comfortably seated, under comfortable room temperature. No bathing, drinking alcohol or caffeine, smoking, exercising or eating 30 minutes before taking a measurement.
Notes:
The arm cuff will start to inflate automatically.

Note: Wait 2-3 minutes before taking another blood pressure measurement. Waiting between readings allows the arteries to return to the condition prior to taking the blood pressure measurement.
Always consult your doctor. Self-diagnosis of measurement results and self-treatment are dangerous.
2. Remove the arm cuff.
3. Press the START/STOP button to turn the monitor off.
The monitor automatically stores the measurement in its memory. It will automatically turn off after 2 minutes.
Important:

| Systolic Blood Pressure | Above 135 mmHg |
| Diastolic Blood Pressure | Above 85 mmHg |
This criteria is for home blood pressure measurement.
). If the irregular heartbeats cause the measurement to be invalid, no result is shown. If the irregular heartbeat symbol (
) is shown after you have taken a measurement, repeat the measurement. If the irregular heartbeat symbol (
) is shown frequently, please make your doctor aware of it.
) will appear on the display. Keep still and repeat the measurement.
The monitor automatically stores the last measurement values (blood pressure and pulse rate).
To View the Readings Stored in Memory
The last measurement value is displayed along with the memory symbol.

Notes:

2. Press the START/STOP button to turn the monitor off.
It will automatically turn off after 2 minutes.
To Delete the Values Stored in Memory
Press and hold the START/STOP button for more than 15 seconds.

Error Display | Cause | Remedy |
![]() | Irregular heartbeats are detected. | Remove the arm cuff. Wait 2 - 3 minutes and then take another measurement. Repeat the steps in section 3.3. If this error continues to appear, contact your doctor |
![]() | Movement during measurement. | Carefully read and repeat the steps in section 3.3. |
![]() | The batteries are low. | You should replace them with new ones ahead of time. Refer to section 2.1. |
![]() | The batteries are exhausted. | You should replace them with new ones at once. Refer to section 2.1. |
![]() | Air plug disconnected. | Insert the plug securely. Refer to section 3.1. |
| Arm cuff not applied correctly. | Apply the arm cuff correctly. Refer to section 3.1. | |
| Air is leaking from the arm cuff. | Replace the cuff with the new one. Refer to Chapter 5. | |
![]() | Movement during measurement and the arm cuff has not been inflated sufficiently | Repeat measurement. Remain still and do not talk during measurement. Refer to section 3.3 |
| If “E2” appears repeatedly, inflate the cuff manually until it is 30 to 40 mmHg above your previous measurement result. Refer to section 3.3. | ||
![]() | The arm cuff was inflated above 299 mmHg when inflating the cuff manually. | Do not inflate the cuff above 299 mmHg. Refer to section 3.3. |
![]() | Movement during measurement. | Repeat measurement. Remain still and do not talk during measurement. Refer to section 3.3. |
![]() | Clothing is interfering with the arm cuff. | Remove any clothing interfering with the arm cuff. Refer to section 3.1. |
![]() | Device error. | Contact your local OMRON representative. |
Problem | Cause | Remedy |
| The reading is extremely low (or high). | Arm cuff not applied correctly. | Apply the arm cuff correctly. Refer to section 3.1. |
| Movement or talking during measurement | Remain still and do not talk during measurement. Refer to section 3.3. | |
| Clothing is interfering with the arm cuff. | Remove any clothing interfering with the arm cuff. Refer to section 3.1. | |
| Arm cuff pressure does not rise. | The air tube is not securely connected into the main unit | Make sure that the air tube is connected securely. Refer to section 3.1 |
| Air is leaking from the arm cuff. | Replace the arm cuff with a new one. Refer to Chapter 5 | |
| Arm cuff deflates too soon. | The arm cuff is loose. | Apply the cuff correctly so that it is firmly wrapped around the arm. Refer to section 3.1. |
| Cannot measure or readings are too low or too high | The arm cuff has not been inflated sufficiently. | Inflate the cuff so that it is 30 to 40 mmHg above your previous measurement result. Refer to section 3.3. |
| Nothing happens when you press the buttons. | The batteries are empty. | Replace the batteries with new ones. Refer to section 2.1 |
| The batteries have been inserted incorrectly | Insert the batteries with the correct (+/-) polarity. Refer to section 2.1. | |
| Other problems. |
If this still does not solve the problem, contact your local OMRON representative. | |
To protect your unit from damage, please observe the following: Cause Remedy Notes:

Note: Read and follow the “Correct Disposal of This Product” in the Technical Data Section when disposing of the device and any used accessories or optional parts.
Calibration and Service
1. Unplug the air plug from the air jack.
2. Gently fold the air tube into the arm cuff.

Notes:

Note: Please check with your local OMRON representatives for the appropriate optional part models
Using the Optional AC Adapter

To disconnect the AC adapter, unplug the AC adapter from the electrical outlet first and then remove the AC adapter plug from the main unit.