Omron NEC-101 Compressor Nebulizer

User Manual - Page 2

For NEC-101.

PDF File Manual, 2 pages, Read Online | Download pdf file

Loading ...
Removing condensation from the air tube
If there is moisture or liquid remaining in the air tube, be sure to follow the
procedure below to remove any moisture from within the air tube.
1) Make sure that the air tube is connected to the air connector on the
compressor.
2) Remove the air tube from the nebulizer kit.
3) Turn on the compressor and pump air through the air tube to expel the
moisture.
Changing the air filter
If the air filter has been used for more than 60 days, replace it with a new
one.
1. Remove the air filter cover using a flat screwdriver
as shown on the right.
2. Change the air filter.
Remove the old air filter and insert the new air filter.
3. Put the air filter cover back in place.
Make sure it is attached properly.
Notes:
• Use only OMRON air filters designed for this device. Do not operate without a filter.
• Do not wash or clean the filter. If the air filter becomes wet, replace it. Damp air filters
can cause blockages.
• There is no front or back orientation for the air filters.
• Check that air filters are clean and free of dust before inserting them.
Troubleshooting
In case of any of the below problems occur during use, first check that no
other electrical device is within 30cm.
If the problem persists, please refer to the table below.
Problem Cause Remedy
Nothing happens when the
power switch is turned on.
Is the power plug plugged into
a power outlet?
Check that the plug is inserted
in a power outlet. Unplug then
reinsert the plug if necessary.
No nebulization or low
nebulization rate, when the
power is on.
Is there medication in the
medication tank?
Add the correct amount of
medication to the medication tank.
Is there too much/little
medication in the medication
tank?
Is the vaporizer head missing
or not assembled correctly?
Attach the vaporizer head
correctly.
Is the nebulizer kit assembled
correctly?
Assemble the nebulizer kit
correctly.
Is the nozzle blocked? Make sure that the nozzle is free
of blockages.
Is the nebulizer kit tilted at a
sharp angle?
Make sure that the nebulizer kit is
not tilted at an angle of more than
30 degrees.
Is the air tube connected
correctly?
Make sure that the air tube
is correctly connected to the
compressor and nebulizer kit.
Is the air tube folded or
damaged?
Make sure that the air tube does
not contain kinks.
Is the air tube blocked? Make sure that the air tube is free
of blockages.
Is the air filter dirty? Replace the air filter with a new
one.
The compressor is
abnormally loud.
Is the air filter cover attached
correctly?
Attach the air filter cover correctly.
The compressor is very hot. Is the compressor covered?
Do not cover the compressor with
any type of cover during use.
Are the ventilation slots
blocked?
Do not block the ventilation slots.
Note:
If the suggested remedy does not solve the problem, do not try to repair the device
- the device are not user serviceable. Please contact an authorized OMRON
representative for assistance.
Cleaning and disinfecting
Caution: Handling the Vaporizer Head
• Always wash the vaporizer head after each use.
• Do not use a brush or pin to clean the vaporizer head.
• When disinfecting the parts by boiling, be sure to boil them in plenty of water.
• Do not boil vaporizer head together with objects other than applicable nebulizer accessories.
Cleaning
Clean the parts after each use to remove residual medication. This will avoid
inefficient nebulization or infection.
Washable parts
•Nebulizer Kit, Mask (PVC), Mouthpiece, Nosepiece
Wash them in warm water and mild detergent (neutral detergent). Rinse them
thoroughly with clean hot tap water and allow to air dry in a clean place.
•Vaporizer Head
Wash it with running water.
Non-washable parts
•Compressor, Air Tube (PVC)
Firstly, make sure that the power plug is unplugged from the power outlet. Wipe
clean with a soft cloth moistened with water or mild detergent (neutral detergent).
•Air Filter
Do not wash or clean the air filter. If the air filter becomes wet, replace it.
Damp air filters can cause blockages.
Disinfecting
Disinfect the parts once a week. If the parts are heavily stained, replace
them with new ones.
To select a method for disinfection, refer to the table below.
o
: applicable
X
: not applicable
Disinfecting
method
Parts
Mouthpiece
Nosepiece
Nebulizer Kit
Adult Mask (PVC)
Child Mask (PVC)
Air Tube
(PVC, 100cm)
Air Filter
Materials
PP
PP
PP
Mask: PVC
(Phthalate free)
Band: Rubber***
PVC
(Phthalate free)
Polyester
A
Alcohol
o o o o
**
X X
Disinfecting
ethanol
Sodium
hypochlorite
o o o o
**
X X
Milton*
(0.1%, 15min.)
Quaternary
ammonium
o o o o
**
X X
Osvan*
(0.1%, 10min.)
Chlorhexidine
o o o o
**
X X
Hibitane*
(0.5%, 30min.)
Amphoteric
Surfactant
o o o o
**
X X
Tego 51*
(0.2%, 15min.)
B
Boiling
o o o
X X X
* an example of commercially available disinfectant. (The concentration and
residence time specified in the table are under the conditions where the service
life of the parts is tested with each disinfectant used as described in its instruction
manual. Please note that the testing was not performed for the purpose of ensuring
the effectiveness of the disinfectants. There is no intention of suggesting that such
disinfectants be used. Conditions of use and ingredients of disinfectant vary for
manufacturers. Please read the instruction manuals carefully before use and apply
disinfectant to each part in an appropriate way. Please notice that service life of the
parts may be shorter depending on conditions, environments and frequency of use.)
** Remove the rubber band from mask before disinfecting.
*** The band not made with natural rubber-latex.
A. Use a commercially available disinfectant. Follow the instructions
provided by the disinfectant manufacturer.
Note: Never clean with benzene, thinner or a flammable
chemical.
B. Parts may be boiled between 15 to 20 minutes.
After boiling, carefully remove the parts, shake off excess
water and allow to air dry in a clean environment.
How to use
Warning:
Clean and disinfect the nebulizer kit, mask, mouthpiece or nosepiece before
using them when as follows:
- the first time after purchase
- if the device has not been used for a long period of time
- if more than one person uses the same device
For the instructions, refer to “Cleaning and disinfecting” section.
1.
Make sure that the power switch is in the off ( )
position.
2.
Plug the power plug into a power outlet.
Note:
Do not place the device in a location where it is difficult to disconnect
the power cord.
3.
Remove the inhalation top from the medication tank.
1) Rotate the inhalation top anti-clockwise.
2) Lift the inhalation top out of the medication tank.
1
2
4.
Add the correct amount of prescribed medication to the
medication tank.
5.
Make sure that the vaporizer head is placed in the
medication tank.
6.
Put the inhalation top back onto the medication tank.
1) Lower the inhalation top on the medication tank.
2) Rotate the inhalation top clockwise.
1
2
7.
Attach the mask, mouthpiece or nosepiece to the nebulizer
kit tightly.
8.
Attach the air tube. Twist the air tube plug slightly and
push it firmly into the air tube connector.
9.
Hold the nebulizer kit as indicated below.
Follow the instructions of your doctor or respiratory therapist.
Caution:
Do not tilt the nebulizer kit at an angle of greater than 30 degrees in all
directions. Medication may flow into the mouth or it may result in not
nebulizing effectively.
Right angle
10.
Turn the power switch to the on ( ) position. As the
compressor starts, nebulization begins and aerosol is
generated.
Inhale the medication.
11.
When treatment is complete, turn the power off and
unplug the compressor from the power outlet.
Important information regarding Electro Magnetic Compatibility (EMC)
NE-C101 manufactured by OMRON HEALTHCARE Co., Ltd. conforms to
EN60601-1-2:2015 Electro Magnetic Compatibility (EMC) standard.
Further documentation in accordance with this EMC standard is available at
http://www.omronhealthcare-ap.com/emc-information.
Refer to the EMC information for NE-C101 on the web site.