This instrument is covered by a 5 year guarantee from the
date of purchase, batteries and accessories are not included.
The guarantee is valid only on presentation of the guarantee
card completed by the dealer conrming date of purchase or
the receipt. Opening or altering the instrument invalidates the
guarantee. The guarantee does not cover damage, accidents
or non-compliance with the instruction manual. Please contact
Rossmax Service.
Specication
Temperature measurement
range
34.0~42.2°C(93.2~108.0°F)
Accuracy ± 0.2°C (0.4°F) during 35.0~42.2°C
(95~108.0°F)
±0.3°C (0.5°F) for other range
Operating temperature
range
10 ~ 40°C (50~104°F), RH<=95%
Storage and transportation
temperature
Temperature:-25~55°C (-13~131°F),
RH <=95%
Memory 9 set
Display resolution 0.1
Battery 3V, lithium CR2032 x 1
Weight (with battery) 55g
Size 12.0cm(L) x 3.5cm(w) x 2.8cm(H)
Probe cover 11 pcs
Auto shut down 60 sec.
Battery life 3000 consecutive measurements or 1
year with 1-2 measurements per day
including stand-by mode.
Safety classication
Type BF equipment
Clinical repeatability 0.08°C (< 1 year old)
0.10°C (1~5 years old)
0.07°C (> 5 years old)
*Dispose of device and batteries according to local regulations
1. Besides the oral temperature, the body temperature measured
by ear may be different from measurements taken by rectal or
axillary. Therefore, there is no any meaning to compare them.
Take the temperature periodically to nd out the normal ear
temperature reading , then use that reading as the basis for
comparison with any measurement taken when you suspect
a fever.
2. This thermometer is designed for ear temperature measurement.
Do not use this thermometer for other body site measurement.
3. Keep the unit dry and away from where it might be exposed to
moisture, liquids, direct sunlight, high temperature, high humidity,
or excessive dust.
4. This device is not shock-protected. Do not drop, heavily pressure
the unit or fall from high place.
5. Do not bend the device.
6. Do not disassemble or make modications on the device.
7. Please do not dispose of the product in the household waste at
the end of its useful life. Disposal can take place at your local
retailer or at appropriate collection points provided in your country.
8. Do not boil the probe.
9. Do not use the device if it operates abnormally or error massage
is showed.
10. Do not use thinner or benzene to clean the device.
11. Wipe the device clean before storing.
12. When taking the device from storage below or above 10~
40°C(50~104°F), place it in 10~ 40°C(50~104°F) temperature
range for at least 30 minutes before use.
13. Remove the battery if this device is not used for a period of time.
14. If this device is used according to the operation instruction, periodic
re-calibration is not required. If you still have questions, please send
the complete device to dealers.
15. Please note that this is a home healthcare product only, and it is
not intended to serve as a substitute for the advice of a physician
or medical professional.
16. Do not use this device for diagnosis or treatment of any health
problem on disease. Measurement results are for reference
only. Contact your physician if you have or suspect any medical
problems. Do not change your medications without the advice of
your physician or healthcare professional.
17. This device may not meet its performance specication if stored
or used outside temperature and humidity ranges specied in
specications.
18. Battery should neither be charged nor placed into the extreme
environment, or it may explode.
19. The thermometer contains small parts (ex: battery, etc.) that can
be swallowed by children. Therefore never leave the thermometer
unattended to children.
20. After wearing the probe cover, the ear thermometer probe must be
vertically and snugly inserted into the ear channel so as to get the
correct temperature reading.
21. Be sure to consult a doctor if you feel that your health is in poor
condition.
22. Do not judge your health only on the presence or absence of a
fever.
Caution
Warranty
IN0RA6000000000XX
OBM_IB_RA600_EN_ver1303
Product Information
Date of purchase:
Store where purchased:
Price Paid (excl. Tax):
Purchase for:
Low battery: Battery
icon is ashing, still can
be measurable.
Suggest to replace the
battery.
Dead battery: Battery
icon always enabled,
no more measure-
ments are possible.
Replace the battery
Electrical fast
transient/burst IEC
61000-4-4
± 2kV for power
supply lines
± 1kV for input/
output lines
Not applicable
Not applicable
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge IEC 61000-4-5 ± 1kV line(s) to
line(s)
± 2kV line(s) to
earth
Not applicable
Not applicable
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage Dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
<5% UT(>95% dip
in UT) for 0,5 cycle
40% UT(60% dip
in UT) for 5 cycles
70% UT(30% dip
in UT) for 25 cycles
<5% UT(>95% dip
in UT) for 5 s
Not applicable
Not applicable
Not applicable
Not applicable
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the RA600 series requires
continued operation during
power mains interruptions, it is
recommended that the RA600
series be powered from an
uninterruptible power supply or
a battery.
Power frequency
(50/60 Hz) magnetic
eld IEC 61000-4-8
3 A/m 3 A/m The RA600 power frequency
magnetic elds should be at
levels characteristic of a typical
location in a typical commercial
or hospital environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Recommended separation distance between portable and mobile RF communications equip-
ment and the RA600
The RA600 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the RA600 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the RA600 as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter / W
Separation distance according to frequency of transmitter / m
150 kHz to 80
MHz / d=1,2√P
80 MHz to 800 MHz
/ d=1,2√P
800 MHz to 2,5 GHz /
d=2,3√P
0,01 N/A 0,12 0,23
0,1 N/A 0,38 0,73
1 N/A 1,2 2,3
10 N/A 3,8 7,3
100 N/A 12 23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reection from structures, objects and people.
Guidance and manufacturer’s declaration-electromagnetic emissions
The RA600 is intended for use in the electromagnetic environment specied below.
The customer or the user of the RA600 should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment-guidance
RF emissions CISPR 11 Group 1 The RA600 uses RF energy only for its internal func-
tion. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby elec-
tronic equipment.
RF emissions CISPR 11 Class B The RA600 is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Not appli-
cable
Voltage uctuations/icker
emissions IEC 61000-3-3
Not appli-
cable
Guidance and manufacturer’s declaration-electromagnetic immunity
The RA600 is intended for use in the electromagnetic environment specied below.
The customer or the user of the RA600 should assure that it is used in such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic environment-guidance
Conducted
RF IEC
61000-4-6
Radiated RF
IEC 61000-
4-3
3 Vrms
150 KHz to
80 MHz
3 V/m
80MHz to
2,5 GHz
Not appli-
cable
3 V/m
Portable and mobile RF communications equip-
ment should be used no closer to any part of the
RA600 series, including cables, than the recom-
mended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance:
d = 1,2 √P
d = 1,2 √P 80MHz to 800 MHz
d = 2,3 √P 800MHz to 2,5 GHz
Where P is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recom-
mended separation distance in metres (m).
Field strengths from xed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of equip-
ment marked with the following symbol:
NOTE1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2: These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reection from structures, objects and people.
a. Field strengths from xed transmitters, such as base stations for radio (cellular/cord-
less) telephones and land mobile radios, amateur radio, AM and FM radio broadcast
and TV broadcast cannot be predicted theoretically with accuracy. To assess the elec-
tromagnetic environment due to xed RF transmitters, an electromagnetic site survey
should be considered. If the measured eld strength in the location in which the
RA600 series is used exceeds the applicable RF compliance level above, the RA600
series should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating
the RA600 series.
b. Over the frequency range 150 kHz to 80 MHz, eld strengths should be less than
3 V/m.
Guidance and manufacturer’s declaration-electromagnetic immunity
The RA600 is intended for use in the electromagnetic environment specied below.
The customer or the user of the RA600 should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance
level
Electromagnetic environment-
guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete
or ceramic tile. If oors are
covered with synthetic material,
the relative humidity should be
at least 30%
EMC guidance and manufacturer’s declaration