Philips C272P4QPKEW/00 Lcd Monitor With Clinical Dimage

User Manual - Page 3

For C272P4QPKEW/00.

PDF File Manual, 63 pages, Read Online | Download pdf file

C272P4QPKEW/00 photo
Loading ...
Loading ...
1
1
. Importan
t
1. Important
The monitor is intended for use with Medical
Equipment to display alpha, numerical and
graphical data. The subject equipment Philips
monitor is powered by an external recognized
AC/DC adaptor. (IEC/EN60601-1).
1.1 Power adapter safety
Information
Power adapter
This adapter (Manufacture: Philips, Model:
PMP60-13-1-HJ-S) is a forming part of the
monitor.
The connection of external equipments
External equipment intended for connection to
signal input/output or other connectors, shall
comply with relevant UL / IEC standard (e.g.
UL 60950 for IT equipment, UL 60601-1 and
ANSI/AAMI ES60601-1 / IEC 60601 series for
systems – shall comply with the standard IEC
60601-1-1, Safety requirements for medical
electrical systems.
Disconnecting Device
The mains plug or appliance coupler is used as
the disconnect device, the disconnect device
shall remain readily operable. Always completely
disconnect the power cord set from your
product whenever you are working or cleaning
on it. Do not make connections while the
power is on, because a sudden rush of power
can damage sensitive electronic components.
&ODVVLÀFDWLRQ
 Degree of protection against the ingress of
water: IPX0
 Equipment not suitable for use in the
presence of a flammable anesthetic mixture
with air or with oxygen or nitrous oxide.
(Non AP or APG Category)
 Mode of operation: Continuous
 Type of protection against electric shock:
Class I ME Equipment
 No Applied part.
Shutdown Procedure
We strongly recommended that you should
shut down the system before you start to clean
any single components.
Please follow the steps below.
 Close all application programs
 Close operating software
 Turn off power switch
 Disconnect power cord set
 Remove all devices
Safety Symbol Description
The following safety symbols are the further
explanations for your reference.
WLWKUHVSHFWWRHOHFWULFVKRFNÀUHDQG
mechanical hazards only in accordance
with ANSI/AAMI ES60601-1, and CAN/
CSA C22.2 NO. 60601-1
Attention, consult ACCOMPANYING
DOCUMENTS.
Type of Current- AC
Dircet Current
European Community Approval,
European Low Voltage Directives
2006/95/EC; European Electromagnetic
Compatibility Directive 2004/108/EC.
TUV Type Testing Approval,
The monitor complies with the
EN60601-1 and IEC60601-1 of
European Standards.
Power "ON"
Power "OFF"
Medical Equipment With respect to
HOHFWULFVKRFNÀUHDQGPHFKDQLFDO
hazards only in accordance with ANSI/
AAMI ES 60601-1:2005, and CAN/
CSA C22.2 NO.60601-1:2008
Note
 Caution: Use suitable mounting apparatus
to avoid risk of injury.
 Use a power cord that matches the voltage
of the power outlet, which has been
Loading ...
Loading ...