KBF / KBF-UL + KMF (E6) 04/2020 page 136/163
22.5 Disposal of the chamber in non-member states of the EU
NOTICE
Danger of violation against existing law if not disposed of properly.
Alteration of the environment.
For final decommissioning and disposal of the chamber, please contact BINDER ser-
vice.
Follow the statutory regulations for appropriate, environmentally friendly disposal.
The main board of the constant climate chamber includes a lithium cell. Please dispose of it according to
national regulations.
The refrigerant used R 134A (1,1,1,2-tetrafluorethane) is not inflammable at ambient pressure. It must not
escape into the environment. In Europe, recovery of the refrigerant R 134A (1300) is mandatory accord-
ing to regulation No. 842/2006/EC. Ensure the compliance with the applicable legal requirements regard-
ing qualification of staff, disposal, and documentation.