TECHNICAL DATA
Fluke 370 FC Series True-rms
Wireless AC/DC Clamp Meters
The Fluke 370 FC Series (376 FC, 375 FC and 374
FC) offers advanced troubleshooting performance.
The iFlex
™
flexible current probe allows easy connections to wires
in tight spaces. In addition, all three clamps are now part of the
Fluke Connect™ family of wireless test tools. Now you can:
• Log and trend measurements to pinpoint intermittent faults
• Transmit results wirelessly via Fluke Connect Measurements app
• Create and send reports right from the field
• Capture measurements outside the arc flash zone with Bluetooth
connectivity to your Apple or Android devices
Product highlights
• Connect your meter to your smartphone using Fluke Connect
Measurements app to increase the troubleshooting power of
your clamp
• Read measurements on your phone at a safe distance, wearing
less PPE while your meter takes all the risks
• Record results directly to your phone and the Fluke Cloud™
• Capture intermittent faults while performing other tasks using the
logging capabilities of the Fluke 376 FC and 375 FC
• Create and share reports from the field via email, or converse in
real time with ShareLive™ video calls
• iFlex flexible current probe expands the measurement range to
2500 A ac; provides access to large conductors in tight spaces
(included with 376 FC; compatible with 375 FC and 374 FC)
• TPAK magnetic hanging strap (included with 376 FC; compatible
with 375 FC and 374 FC) allows you to conveniently hang the
meter while using the probes
• Integrated VFD low pass filter (Fluke 376 FC and 375 FC) for
accurate motor drive measurements
• Proprietary in-rush measurement technology to filter out
noise and capture motor starting current exactly as the circuit
protection sees it
• CAT IV 600 V, CAT III 1000 V safety rating
• Three-year warranty
• Soft carrying case
LOG AND TREND MEASUREMENTS
Transmit results wirelessly via Fluke Connect
Measurements app
WORK SAFELY
Capture measurements outside the arc flash
zone with Bluetooth connectivity to your Apple or
Android devices
LO-PASS FILTER
Integrated VFD low-pass filter
(Fluke 376 FC and 375 FC) for accurate motor
drive measurements
TRUE-RMS
True-rms voltage and current for accurate mea-
surements on non-linear signals
WARRANTY
Three-year

2 Fluke Corporation Fluke 370 FC Series True-rms Wireless AC/DC Clamp Meters
Comparison chart
Fluke 376 FC Fluke 375 FC Fluke 374 FC
CAT rating CAT III 1000 V, CAT IV 600 V CAT III 1000 V, CAT IV 600 V CAT III 1000 V, CAT IV 600 V
A ac range 999.9 A with clamp jaws
2500 A with iFlex flexible
current clamp
600 A with clamp jaws
2500 A with optional iFlex
flexible current clamp
600 A with clamp jaws
2500 A with optional iFlex
flexible current clamp
A dc range 999.9 A 600 A 600A
Best resolution 0.1 A 0.1 A 0.1 A
V ac range 1000 V 1000 V 1000 V
V dc range 1000 V 1000 V 1000 V
mV dc range 500 mV 500 mV
Ohms range 60.00 kΩ 60.00 kΩ 6000 Ω
Frequency 5-500.0 Hz 5-500.0 Hz
Capacitance 1 μF-1,000 μF 1 μF-1,000 μF 1 μ-1,000 μF
Auto shutoff
• • •
True RMS
• • •
Display hold
• • •
Inrush measurements
• • •
Backlight
• • •
DC zero
• • •
Min/max
• • •
VFD low pass filter
• •
-
Logging
• •
-
iFlex flexible current clamp Included Optional Optional
Fluke Connect compatible
• • •
Measurement capability
• 1000 A ac and dc current measurement (376 FC); 600 A ac and dc (375 FC and 364 FC)
• 2500 A ac current measurement with iFlex® flexible current probe
• 1000 V ac and dc voltage measurement
• True-rms voltage and current for accurate measurements
on non-linear signals
• Frequency measurement to 500 Hz with body jaw and iFlex (376 FC and 375 FC)
• Resistance measurement to 60 kΩ (376 FC and 375 FC) or 6000 Ω (374 FC), with continuity detection
• Min, max, average and in-rush recording to capture variations automatically
• 500 mV dc measurement range to interface with other accessories (376 FC and 375 FC)
• 1000 μF capacitance measurement

3 Fluke Corporation Fluke 370 FC Series True-rms Wireless AC/DC Clamp Meters
Specifications
General specifications
Maximum voltage
(between any terminal and earth ground)
1000 V
Battery type 2 AA alkaline batteries, IEC LR6
Temperature
Operating -10 °C to +50 °C (14 °F to + 122 °F)
Storage -40 °C to +60 °C (-40 °F to + 140 °F)
Operating humidity Non condensing (< 10 °C (50 °F)), ≤90 % RH (at 10 °C to 30 °C (50 °F to 86 °F)), ≤75 % RH
(at 30 °C to 40 °C (86 °F to 104 °F)), ≤45 % RH (at 40 °C to 50 °C (104 °F to 122 °F))
Altitude
Operating 3000 m (9842 ft)
Storage 12,000 m (39,370 ft)
Dimensions 24.9 cm (9.8 in) x 8.5 cm (3.3 in) x 4.5 cm (1.7 in)
Weight 395 g (13.9 oz)
Jaw opening 34 mm (1.3 in)
Flexible current probe diameter 7.5 mm (.29 in)
Flexible current probe cable length
(head to electronics connector)
1.8 m (70.8 in)
Safety IEC 61010-1, Pollution Degree 2
IEC 61010-2-032: CAT III 1000 V / CAT IV 600 V
IEC 61010-2-033: CAT III 1000 V / CAT IV 600 V
IP rating IEC 60529: IP30, non-operating
Radio frequency certification FCC ID:T68-FBLE IC:6627A-FBLE
Electromagnetic Compatibility (EMC) International: IEC 61326-1: Portable, Electromagnetic Environment,
IEC 61326-2-2 CISPR 11: Group 1, Class A
Korea (KCC): Class A Equipment (Industrial Broadcasting & Communication Equipment)
USA (FCC): 47 CFR 15 subpart B. This product is considered an exempt device per clause
15.103.
Temperature coefficients Add 0.1 × specified accuracy for each degree C above 28 °C or below 18 °C

4 Fluke Corporation Fluke 370 FC Series True-rms Wireless AC/DC Clamp Meters
Specifications (continued)
Function Range Resolution Accuracy Crest factor Trigger level
AC current
via jaw
374 FC and 375 FC:
600.0 A
376 FC: 999.9 A
0.1 A
2 % ±5 digits
(10 Hz to 100 Hz)
3 at 500 A
(375 FC and 376 FC only)
2.5 at 600 A
1.42 at 1000 A
(376 FC only)
Add 2 % for C.F. >2
-
AC current via
flexible current
probe
2500 A
374 FC and 375 FC:
0.1 A (≤ 600.0 A),
1 A (≤ 2500 A)
376 FC:
0.1 A (≤ 999.9 A),
1 A (≤ 2500 A)
3 % ±5 digits
(5 Hz to 500 Hz)
3.0 at 1100 A
(375 FC and 376 FC only)
2.5 at 1400 A
1.42 at 2500 A
Add 2 % for C.F. >2
-
DC current
374 FC and 375 FC:
600.0 A
376 FC: 999.9 A
0.1 A 2 % ±5 digits - -
AC voltage 1000 V
0.1 V (≤600.0 V)
1 V (≤1000 V)
1 % ±5 digits
(20 Hz to 500 Hz)
- -
DC voltage 1000 V
0.1 V (≤600.0 V)
1 V (≤1000 V)
1 % ±5 digits - -
mV dc 500.0 mV 0.1 mV 1 % ±5 digits - -
Frequency
via jaw
375 FC and 376 FC:
5.0 Hz to 500.0 Hz
0.1 Hz 0.5 % ±5 digits -
5 Hz to 10 Hz, ≥10 A
10 Hz to 100 Hz, ≥5 A
100 Hz to 500 Hz, ≥10 A
Frequency via
flexible current
probe
375 FC and 376 FC:
5.0 Hz to 500.0 Hz
0.1 Hz 0.5 % ±5 digits -
5 Hz to 20 Hz, ≥25 A
20 Hz to 100 Hz, ≥20 A
100 Hz to 500 Hz, ≥25 A
Resistance
374 FC: 6,000 Ω
375 FC and 376 FC:
60 kΩ
374 FC:
0.1 Ω (≤600 Ω),
1 Ω (≤6000 Ω)
375 FC and 376 FC:
0.1 Ω (≤600 Ω),
1 Ω (≤6000 Ω),
10 Ω (≤60 kΩ)
1 % ±5 digits - -
Capacitance 1000 μF
0.1 μF (≤100 μF),
1 μ F (≤1000 μF)
1 % ±4 digits - -

5 Fluke Corporation Fluke 370 FC Series True-rms Wireless AC/DC Clamp Meters
Ordering information
374 FC 600A AC/DC True-rms Wireless Clamp
Meter
375 FC 600A AC/DC True-rms Wireless Clamp
Meter
376 FC 1000A AC/DC True-rms Wireless
Clamp Meter with iFlex
Included
18-inch iFlex Flexible Current Probe
(376 FC only)
TPAK Magnetic Hanging Strap (376 FC only)
Soft Carrying Case
TL75 Test Leads
Two AA Alkaline Batteries
Coated Instruction Card
Safety Information Sheet
Optional accessories
TPAK Magnetic Hanging Strap
(included with 376 FC)
18-inch iFlex Flexible Current Probe
(included with 376 FC)
Visit www.fluke.com to get complete details
on these products or ask your local Fluke
sales representative.
Fluke Corporation
PO Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
Fluke. Keeping your world
up and running.
®
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0)40 267 5100 or
Fax +31 (0)40 267 5222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2015, 2020 Fluke Corporation.
Specifications subject to change without notice.
3/2020 6005995c-en
Modification of this document is not
permitted without written permission
from Fluke Corporation.
All trademarks are the property of their respective owners. WiFi or cellular service required to share
data. Smartphone, wireless service and data plan not included with purchase. First 5GB of storage is
free.
Smartphone wireless service and data plan not included with purchase. Fluke Connect
is not available in all countries.
Preventive maintenance simplified. Rework eliminated.
Save time and improve the reliability of your maintenance data by
wirelessly syncing measurements using the Fluke Connect system.
• Eliminate data-entry errors by saving measurements directly from the
tool and associating them with the work order, report or asset record.
• Maximize uptime and make confident maintenance decisions with data
you can trust and trace.
• Move away from clipboards, notebooks and multiple spreadsheets with
a wireless one-step measurement transfer.
• Access baseline, historical and current measurements by asset.
• Share your measurement data using ShareLive™ video calls and emails.
• The Ti300+ infrared camera is part of a growing system of connected
test tools and equipment maintenance software. Visit the Fluke website
to learn more about the Fluke Connect system.
Find out more at fluke.com