Operation
Procedure for Generating System Pressure 3
3-5
Procedure for Generating System Pressure
TEST PORT
INTENSIFIER
RESERVOIR
VALVE (3)
LP VALVE (2)
HP VALVE (1)
CAPSTAN
PCU
glf07.eps
Figure 3-6. Generate System Pressure
1. Close HP valve (1).
2. Open LP valve (2).
3. Turn capstan fully out to prime it with fluid from the reservoir.
1.888.610.7664 sales@GlobalTestSupply.com