Maintaining the Test Tool
Using and Adjusting 10:1 Scope Probes
2
47
Note
Ensure that the battery pack is placed in the
battery compartment as shown in Figure 2-2.
8. Reinstall the battery cover and secure the screw.
Using and Adjusting 10:1 Scope Probes
Note
The 10:1 voltage probe VPS40 that is supplied
with Fluke 124 is always adjusted correctly to the
Test Tool and needs no further adjustment.
You need to adjust other scope probes for optimal
response.
Warning
To avoid electrical shock use the BB120
Banana-to-BNC adapter (delivered with the
test tool) to connect a 10:1 scope probe to the
input of the test tool.
To adjust probes, do the following:
Connect the 10:1 scope probe from the gray input B
jack to the red input A jack. Use the red 4-mm banana
adapter (delivered with the probe) and the banana-to-
BNC adapter (BB120). See Figure 2-3.
BB120
Figure 2-3. Adjusting Scope Probes