1-4
Temperature and Air Density variations are less significant than gravity.
Variations should be corrected for when maximum accuracy is required.
Temperature variation example:
Deadweight Tester calibrated temperature 20 °C
Operating temperature 24 °C
Percentage change per °C 0.002%
Indicated Pressure 250 psi
()
250
100
002.0
2420250 xxPRESSURETRUE +=
psiPRESSURETRUE 98.249=
Model Variations
This manual covers many pressure ranges and models within the P3000 series of
instruments.
The following illustrations detail the physical appearance of the six basic models: