T6-1000/T6-600
Service Information
8
6.
Verify that the Product reading is within the display limits.
7.
Apply the test values for the remaining steps of Table 7. Verify the readings are within the display limits.
Table 7. FieldSense Low Current Tests
8.
Set the calibrator to STANDBY. Change the connection from the calibrator AUX HI to the calibrator AUX 20A jack.
9.
Set the calibrator to the values listed in Table 8, Step 1, and apply to the 10/50-turn coil.
10.
Verify that the Tester reading is within the display limits.
11.
Apply the test values shown in the remaining steps of Table 8 and verify the Tester readings are within the display limits.
Table 8. FieldSense High Current Tests
Step
Calibrator Output
(AUX HI)
T6-600/T6-1000 T6-1000 only
Display Limits Display Limits
1 0.5 A, 45 Hz
4.6 A to 5.4 A 43 Hz to 47 Hz
2 0.5 A, 60 Hz
4.6 A to 5.4 A 58 Hz to 62 Hz
3 0.5 A, 66 Hz
4.6 A to 5.4 A 64 Hz to 68 Hz
Step Calibrator Output Display Limits
1 10 A, 45 Hz
96.7 A to 103.3 A
2 10 A, 60 Hz
96.7 A to 103.3 A
3 10 A, 66 Hz
96.7 A to 103.3 A
4 18 A, 45 Hz
174.3 A to 185.7 A
5 18 A, 60 Hz
174.3 A to 185.7 A
6 18 A, 66 Hz
174.3 A to 185.7 A