Loading ...
Loading ...
Loading ...
Clamp Meter
Temperature Zero Procedure (324 and 325)
19
Temperature Zero Procedure (324 and 325)
1. Turn off the Product and wait 20 minutes.
2. Connect a K-type therm
ocouple connector from the Calibrator to th
e Product.
3. After 20 m
inutes, with the Product still connected to the 3.0 V, turn on th
e Product
and short across the CAL keypad on
th
e pca. See Figure 6.
4. Turn the Product rotary switch to tem
per
ature.
5. Push twice until “C-23” shows on t
he Product display.
6. Apply 0 °C from
the Calibrator.
7. Push the HOLD button
to confirm C-23
.
8. Push the HOLD button
to save the calib
ration data.
The calibration adjustm
ent procedure is complete.
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com