323/324/325
Users Manual
12
Emissions that exceed the levels required by CISPR 11 can occur
when the equipment is connected to a test object.
Korea (KCC) Class A Equipment (Industrial Broadcasting & Communication Equipment)
Class A: Equipment meets requirements for industrial electromagnetic
wave equipment and the seller or user should take notice of it. This
equipment is intended for use in business environments and not to be
used in homes.
USA (FCC) 47 CFR 15 subpart B. This product is considered an exempt device per
clause 15.103.
Temperature Coefficients ........................ Add 0.1 x specified accuracy for each degree C
above 28 °C or below 18 °C
Resolution
323 .................................................... 0.1 A
324, 325 ............................................ (0.01, 0.1) A
Accuracy
323, 325 ............................................ 2.0 % ±5 digits (45 – 65 Hz)
2.5 % ±5 digits (65 – 400 Hz)
324 .................................................... 1.5 % ±5 digits (45 Hz to 400 Hz)
Note
Add 2 % for position sensitivity.