Use the following procedure to make a measurement:
1. Connect the i1000s Current Probe to the desired input channel on the
oscilloscope.
2. On the Current Probe, select the least sensitive range (1 mV/A). See
Figure 1 for selector switch location.
3. On your oscilloscope, select an appropriate range. Use a 1:1 probe setting.
Note
If using the Fluke ScopeMeter 123, use this procedure for step 3:
1. Press
(shown in Figure 3).
2. When the MEASUREMENTS screen appears, select AMP…
and ENTER.
3. When ENTER is pressed, the AMPERE PROBE screen
appears.
4. Select the appropriate range (i.e. 1 mV/A).
4. Clamp the probe around the conductor to be measured, and observe the
current waveform on your oscilloscope display.
Note
If using the Fluke 123 ScopeMeter, disregard the next step.
5. Calculate the Amps-per-division on your oscilloscope by dividing the
vertical scale (mV or V per division) by the i1000s switch position
(1, 10, or 100 mV/A). For example:
20 mV per division = 20 A per division
1 mV/A
Setups using the Current Probe and a ScopeMeter are shown in Figure 3.