Digital Multimeter
How to Make Measurements
15
Resistance Measurements
XW Warning
To prevent possible electrical shock, fire, or
personal injury, disconnect power and
discharge all high-voltage capacitors before
you measure resistance, continuity,
capacitance, or a diode junction.
The Meter measures resistance by sending a small
current through the circuit. Because this current flows
through all possible paths between the probes, the
resistance reading represents the total resistance of all
paths between the probes.
The Meter's resistance ranges are 600.0 Ω, 6.000 kΩ,
60.00 kΩ, 600.0 kΩ, 6.000 MΩ, and 50.00 MΩ.
Set up the Meter as shown in Figure 5 to measure
resistance.
The following are some tips for measuring resistance:
The measured value of a resistor in a circuit is often
different from the resistor's rated value.
The test leads can add 0.1 Ω to 0.2 Ω of error to
resistance measurements. To test the leads, touch
the probe tips together and read the resistance of the
leads. If necessary, you can use the relative (REL)
mode to automatically subtract this value.
The resistance function can produce enough voltage
to forward-bias silicon diode or transistor junctions,
causing them to conduct. If this is suspected, press
to apply a lower current in the next higher
range. If the value is higher, use the higher value.
Refer to the Input Characteristics table in the
specifications section for typical short-circuit currents.