381
Calibration Manual
8
Required Equipment
The equipment listed in Table 2 is required for performance tests and calibration
adjustment.
Table 2. Required Equipment
Equipment Required Characteristics Recommended Model
Calibrator 4.5-digit resolution Fluke 552xA Calibrator
Wired coil 50 turns 5500A/COIL
Test Probe for iFlex 2 mm to 4 mm Slim reach probe TP2, PN650892
Test Lead Test Lead w/retractable sheath 6358, PN1903307
Performance Tests
XWWarning
To prevent possible electrical shock, fire, or personal injury,
do not perform the performance test procedures unless the
Product is fully assembled.
The following performance tests verify the complete operation of the Product and check
the accuracy of each function against the specifications of the Product. See Table 3. If the
Product fails any part of the test, calibration adjustment and/or repair is indicated. See
“Calibration Adjustment”. Table 3 shows the Performance tests for the product. Table 4
shows the Performance tests for the iFlex. Figures 3 and 4 show the performance test
connections for ac amps and dc amps. Figure 5 shows the performance test connections
for ac volts, dc volts, and frequency.
0
/
+
REMOTE DISPLAY
T
RMS
CLAMP ME TER
381
iFlex
ghn10.eps
Figure 3. Performance Test Connections for AC Amps and DC Amps