70
DC Voltage Specification
Accuracy
AC over DC, DC over AC, AC + DC
[2]
Function Range Resolution
DC
[2]
20 to 45 Hz 45 Hz to 1 kHz 1 to 20 kHz 20 to 35 kHz
50 mV
[3]
0.001 mV 0.05 % + 20
[4]
1.5 % + 40 5 % + 40 DC mV
500 mV 0.01 mV 0.025 % + 2
[5]
1.5 % + 40 5 % + 40
5 V 0.0001 V 0.025 % + 2 1.5 % + 40 5 % + 40
50 V 0.001 V 0.025 % + 2 1.5 % + 40 5 % + 40
500 V 0.01 V 0.03 % + 2 Not Spec’d Not Spec’d
DC V
[1]
1000 V 0.1 V 0.03 % + 2
2 % + 80 0.5 % + 80
Not Spec’d Not Spec’d
L
[1]
1000 V 0.1 V 1 % + 20 Not Spec’d Not Spec’d Not Spec’d Not Spec’d
[1] Add 20 counts in dual display ac over dc, dc over ac or ac+dc.
[2] AC+DC ranges are specified from 2 % to 140 % of range, except 1000 V is specified from 2 % to 100 % of range.
[3] When using the relative mode (REL Q) to compensate for offsets.
[4] Add 4 counts/10 mV AC in dual display ac over dc, dc over ac or ac + dc
[5] Add 10 counts/100 mV AC in dual display ac over dc, dc over ac or ac + dc.