Summit ACR1322GNSF456 24" Wide Upright Healthcare Refrigerator, Certified to NSF/ANSI 456 Vaccine Storage Standard

User Manual - Page 8

For ACR1322GNSF456.

PDF File Manual, 15 pages, Read Online | Download pdf file

ACR1322GNSF456 photo
Loading ...
Loading ...
Loading ...
7
OPERATING YOUR APPLIANCE
Temperature Controller
Each refrigerator has a factory preset temperature controller (thermostat). The default setpoint for your
unit is 4°C (39°F). To confirm or change the temperature setting, follow the steps below:
TO SEE THE SET POINT
The default display for your temperature controller is the internal cabinet temperature. To view the unit’s set point, push
and immediately release the SET key. The display will flash, showing the set point value. If no other actions are taken in
6 seconds, the display will revert to the internal air temperature.
Please note the temperature displayed on the temperature controller is the internal air temperature. Your temperature
measurement device/alarm/thermometer (installed on the left in the control panel) provides a temperature readout based
on the buffered probe to better simulate the temperature of what is stored inside the unit (as opposed to the air
temperature itself). As a result, these readings may not always display the same temperature reading.
Changes to the settings are not recommended; any changes may compromise the unit’s
ability to perform in compliance with the NSF/ANSI 456 Standard.
If you find it necessary to change the temperature setting for your temperature controller,
follow the instructions below:
To change the SET POINT
• To prevent accidental adjustment, the temperature controller is locked. To unlock it, press the DOWN arrow and hold
for six seconds and ON will be displayed.
• To change the set point, push and immediately release the SET key. The display will flash, while the current
set point value is blinking, press the UP or DOWN arrow to change the set point to your desired temperature.
Press the SET button again to exit the adjustment. The display will revert to the internal air temperature. To
lock the thermostat, press the down arrow and hold for six seconds and OFF will be displayed.
NOTE: The temperature controller displays the set point after the unit stabilizes to the set temperature. It only
reads the air temperature during the stabilization period.