Pressure Calibrator
Transmitter Calibration
29
Ranges and Resolution
Ranges and resolutions for the Product are shown in Table 4.
Table 4. Ranges and Resolutions
Range (PSI) 16 36 100 300 500 1000 1500 3000 5000
Burst Pressure 60 120 400 1200 2000 4000 6000 9000 10000
Proof Pressure
(PSI)
35 70 200 600 1000 2000 3000 6000 7000
Engineering Unit Factor
psi 1 16 36 100 300 500 1000 1500 3000 5000
bar 0.06894757 1.1032 2.4821 6.8947 20.684 34.474 68.947 103.42 206.84 344.74
mbar 68.94757 1103.2 2482.1 6894.8 20684 34474 68948 N/A N/A N/A
kPa 6.894757 110.32 248.21 689.48 2068.4 3447.4 6894.8 10342 20684 34474
MPa 0.00689476 0.1103 0.2482 0.6894 2.0684 3.4474 6.8948 10.342 20.684 34.474
kg/cm2 0.07030697 1.1249 2.5311 7.0307 21.092 35.153 70.307 105.46 210.92 351.53
cmH20 @ 4 °C
70.3089 1124.9 2531.1 7030.9 21093 35154 70309 N/A N/A N/A
cmH20 @ 20 °C
70.4336 1126.9 2535.6 7043.4 21130 35217 70434 N/A N/A N/A
mmH20 @ 4 °C
703.089 11249 25311 70309 N/A N/A N/A N/A N/A N/A
mmH20 @ 20 °C
704.336 11269 25356 70434 N/A N/A N/A N/A N/A N/A
inH20 @ 4 °C
27.68067 442.89 996.50 2768.1 8304.2 13840 27681 41521 83042 N/A
inH20 @ 20 °C
27.72977 443.68 998.27 2773.0 8318.9 13865 27730 41595 83189 N/A
inH20 @ 60 °F
27.70759 443.32 997.47 2770.8 8312.3 13854 27708 41561 83123 N/A
mmHg @ 0 °C
51.71508 827.44 1861.7 5171.5 15515 25858 51715 77573 N/A N/A
inHg @ 0 °C
2.03602 32.576 73.297 203.60 610.81 1018.0 2036.0 3054.0 6108.1 10180
Proof pressure - maximum allowable pressure without a shift in calibration.
Burst pressure - sensor damaged or destroyed; some risk of personnel injury.