FLUKE 289-NIST True RMS Industrial Data Logging Multimeter, includes Traceable Certificate

User Manual for the Fluke 289-NIST True RMS Industrial Data Logging Multimeter, - Page 77

For 289-NIST.

PDF File Manual, 87 pages, Read Online | Download pdf file

289-NIST photo
Loading ...
Loading ...
Loading ...
287/289
Users Manual
68
AC Voltage Specifications
Accuracy
Function Range Resolution
20 to 45 Hz 45 to 65 Hz 65 Hz to 10 kHz 10 to 20 kHz 20 to 100 kHz
50 mV
[1]
0.001 mV 1.5 % + 60 0.3 % + 25 0.4 % + 25 0.7 % + 40 3.5 % + 40
[5]
AC mV
500 mV 0.01 mV 1.5 % + 60 0.3 % + 25 0.4 % + 25 0.7 % + 40 3.5 % + 40
5 V
[1]
0.0001 V 1.5 % + 60 0.3 % + 25 0.6 % + 25 1.5 % + 40 3.5 % + 40
[5]
50 V
[1]
0.001 V 1.5 % + 60 0.3 % + 25 0.4 % + 25 0.7 % + 40 3.5 % + 40
500 V
[1]
0.01 V 1.5 % + 60 0.3 % + 25 0.4 % + 25 Not Spec’d Not Spec’d
AC V
1000 V 0.1 V 1.5 % + 60 0.3 % + 25 0.4 % + 25 Not Spec’d Not Spec’d
-70 to -62 dB
[3]
0.01 dB 3 dB 1.5 dB 2 dB 2 dB 3 dB
-62 to -52 dB
[3]
0.01 dB 1.5 dB 1.0 dB 1 dB 1 dB 2 dB
-52 to -6 dB
[3]
0.01 dB 0.2 dB 0.1 dB 0.1 dB 0.2 dB 0.8 dB
-6 to +34 dB
[3]
0.01 dB 0.2 dB 0.1 dB 0.1 dB 0.2 dB 0.8 dB
dBV
34 to 60 dB
[3]
0.01 dB 0.2 dB 0.1 dB 0.1 dB Not Spec’d Not Spec’d
Low pass filter
[4]
2 % + 80 2 % + 40
2 % +10
-6 % -60
[2]
Not Spec’d Not Spec’d
L
[4]
1000 V 0.1 V 2 % + 80 2 % + 40 2 % + 40
[6]
Not Spec’d Not Spec’d
[1] Below 5 % of range, add 20 counts.
[2] Specification increases linearly from -2 % at 200 Hz to -6 % at 440 Hz. Range is limited to 440 Hz.
[3] dBm (600 Ω) is specified by adding +2.2 dB to the dBV range values.
[4] 289 only.
[5] Add 2.5 % above 65 kHz.
[6] Range is limited to 440 Hz.
See Detailed Specifications introduction for additional information.