FLUKE 1551A-9-DL Intrinsically Safe Stik Thermometer

Product's Documents

Below are documents related to this product, you can read online or download:
User Manual Specification
1551A-9-DL photo

Instruction Sheet

This is the main product document for model 1551A-9-DL. Additionally, the document applies to other FLUKE models: 1551A EX, 1552A EX

The file format is pdf, 2 pages, you can download this manual here .

background
1551A Ex/1552A Ex English Instruction Sheet
Page 1
November 2010
© 2010 Fluke Corporation. All rights reserved.
All product names are trademarks of their respective companies.
1551A Ex/1552A Ex
Immersion Characteristics
Instruction Sheet
Introduction
The 1551A Ex/1552A Ex Stik Thermometer (Theremometer)
must be adequately immersed in the medium to make an
accurate temperature measurement.
Immersion Depth
The minimum required immersion depth depends on the type of
medium (whether wet or dry) as shown in Table 1. These figures
should result in immersion errors of less than 0.005 °C in most
cases. Actual results can vary depending on the type and form of
the surrounding materials.
Table 1. Immersion Depth
Medium Minimum Immersion Depth
Wet (liquid bath, thermowell with
thermal compound)
7 cm (2.8 in.)
Dry (dry-well, thermowell without
thermal compound)
12 cm (4.8 in.)
Time Constant
The different probe sizes of the Thermometer exhibit different
settling time constants. This parameter describes the time it
takes for the Thermometer to change 63 % toward its final
temperature (see ASTM E 644).The time constant is primarily
dependent on the diameter of the metal sheath of the
Thermometer. Typical approximate time constants are given in
the following table.
Table 2. Time Constant
Sheath Diameter Time Constant
4.8 mm (3/16 in.) 14 seconds
6.35 mm (1/4 in.) 21 seconds