FLUKE 5609-400-B Secondary PRT with Bare Wire, 400 x 6mm, Calibration Options

User Manual for the Fluke 5609-400-B Secondary PRT with Bare Wire, 400 x 6mm, Calibration Options - Page 12

For 5609-400-B.

PDF File Manual, 17 pages, Read Online | Download pdf file

5609-400-B photo
Loading ...
Loading ...
Loading ...
DO allow sufficient time for the thermometer to stabilize before making
measurements. This allows for the best accuracy.
DO use the correct drive current with the thermometer to prevent error
in temperature or resistance. Hart Scientific recommends 1mA.
DO anneal the 5608 thermometer at 500°C and the 5609 thermome-
ter at 600 to 661°C for 12 hours if they become oxidized.
DO use the protective shipping box or case provided or other protec-
tion when the thermometer is not in use.
DO NOT subject the thermometer to any physical shock or vibration.
DO NOT use pliers or other devices to squeeze the sheath. This action
can permanently damage the PRT.
DO NOT subject the thermometer to temperatures above the highest
specified operating temperature.
DO NOT expose the thermometer’s handle or cables to extreme tem-
peratures. The temperature limits of the handle are: –50°C to 200°C
DO NOT submerge the handle or cable in liquids.
DO NOT screw a clamp down so tight that it dents the sheath. This can
permanently damage the PRT.
5608/5609
User’s Guide
12
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com
Loading ...
Loading ...
Loading ...