11
Specications and Environmental Conditions
Specications
2 Specications and Environmental Conditions
2.1 Specications
Table 2 Base Unit Specifications
Base Unit Specifications
9142 9143 9144
Temperature Range at
23°C
–25°C to 150°C
(-13°F to 302°F)
33°C to 350°C
(91°F to 662°F)
50°C to 660°C
(122°F to 1220°F)
Display Accuracy ±0.2°C Full Range ±0.2°C Full Range ±0.35°C at 50°C
±0.35°C at 420°C
±0.5°C at 660°C
Stability ±0.01°C Full Range ±0.02 °C at 33°C
±0.02 °C at 200°C
±0.03°C at 350°C
±0.03°C at 50°C
±0.05°C at 420°C
±0.05°C at 660°C
Axial Uniformity at 40
mm (1.6 in)
±0.05°C Full Range ±0.04°C at 33°C
±0.1°C at 200°C
±0.2°C at 350°C
±0.05°C at 50°C
±0.35°C at 420°C
±0.5°C at 660°C
Axial Uniformity at 60
mm (2.4 in)
±0.07°C Full Range ±0.04°C at 33°C
±0.2°C at 200°C
±0.25°C at 350°C
±0.1°C at 50°C
±0.6°C at 420°C
±0.8°C at 660°C
Radial Uniformity ±0.01°C Full Range ±0.01°C at 33°C
±0.015°C at 200°C
±0.02°C at 350°C
±0.02°C at 50°C
±0.05°C at 420°C
±0.1°C at 660°C
Loading Effect (with
a 6.35 mm reference
probe and three 6.35
mm probes)
±0.006°C Full Range ±0.015°C Full Range ±0.015°C at 50°C
±0.025°C at 420°C
±0.035°C at 660°C
Loading Effect (versus
display with 6.35 mm
probes)
±0.08°C Full Range ±0.2°C Full Range ±0.1°C at 50°C
±0.2°C at 420°C
±0.2°C at 660°C
Hysteresis 0.025°C 0.03°C 0.1°C
Operating Conditions 0°C to 50°C, 0% to 90% RH (non-condensing)
Environmental
conditions for all
specications except
temperature range
13°C to 33°C
Immersion (Well)
Depth
150 mm (5.9 in)
Insert OD 30 mm (1.18 in) 25.3 mm (1.00 in) 24.4 mm (0.96 in)
Heating Time 16 min: 23°C to 140°C
23 min: 23°C to 150°C
25 min: –25°C to 150°C
5 min: 33°C to 350°C 15 min: 50°C to 660°C
Cooling Time 15 min: 23°C to –25°C
25 min: 150°C to –23°C
32 min: 350°C to 33°C
14 min: 350°C to 100°C
35 min: 660°C to 50°C
25 min: 660°C to 100°C
Resolution 0.01°
Display LCD,°C or°F user-selectable
Key Pad Arrows, Menu, Enter, Exit, 4 soft keys
Size (H x W x D) 290 mm x 185 mm x 295 mm (11.4 x 7.3 x 11.6 in)