FLUKE 5627A-12-D Precision Thermometer/RTD Temperature Probe with 5-Pin DIN for Tweener Thermometers, 12 x 0.25in, -200 to 300°C

User Manual for the Fluke 5627A-12-D Precision Thermometer/RTD Temperature Probe with 5-Pin DIN for Tweener Thermometers, 12 x 0.25in, -200 to 300°C - Page 7

For 5627A-12-D.

PDF File Manual, 12 pages, Read Online | Download pdf file

5627A-12-D photo
Loading ...
Loading ...
Loading ...
3 Specifications
Resistance
Nominal 100
Ω
Temperature Coefficient
0.00385
Ω
/
Ω
/ °C nominal
Temperature Range –200 °C to 420 °C
(5627A-6 and 5627A-9 to 300 °C; transition and cable temperature: 0 °C to
150 °C)
Drift Rate
±0.13 °C at 0 °C after 1000 hours at 400 °C
Sheath Material
316 Stainless Steel
Transition junction
temperature range
–50 °C to 125 °C
Leads
Teflon
-insulated, nickel-plated stranded copper, 22 AWG
Time Constant Four seconds maximum for 63.2 % response to step change in water moving
at 3 fps.
Bending Radius Sheath may be ordered with a bend on a minimum radius of 19 mm (3/4 in) ex
-
cept for 50 mm (2 in) area of sheath near tip. (Hart lab requires 20 cm [8 in] of
unbent sheath to re-calibrate.)
Calibration
Includes manufacturer’s NVLAP-accredited calibration
Immersion
At least 100 mm (4 in) recommended
Size 5627A-12:12 in L x 1/4 in Dia.
5627A-9: 9 in L x 3/16 in Dia.
5627A-6: 6 in L x 3/16 in Dia.
Temperatures outside this range will should not be too hot to touch.cause irreparable damage. For best
performance, transition junction
7
3 Specifications
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com
Loading ...
Loading ...
Loading ...